The ventricular fibrillation therapy assurance (VFTA) algorithm in newer Abbott ICDs can malfunction and deliver inappropriate therapies, necessitating its deactivation.
Reported VFTA malfunction raises caution for VF detection in Abbott ICDs; leaves open broader algorithm safety assessment.
Newer devices incorporate advanced algorithms to have more accurate delivery of implantable cardioverter-defibrillator (ICD) therapy. Newer models of Abbott Implantable cardioverter-defibrillators like Gallant and Entrant have a new ventricular fibrillation (VF) therapy assurance algorithm. This algorithm has a proven role in avoiding under-detection of VF where some of the smaller electrograms are under-sensed. We report a case of ventricular fibrillation therapy assurance (VFTA) algorithm malfunction in a 48-year-old lady having single chamber Entrant Implantable cardioverter-defibrillator, where the patient received inappropriate therapies. The VFTA algorithm eventually had to be turned off.