Investigates NIRS measurement accuracy for oxygen consumption and blood flow in subjects, suggesting limited quantification reliability.
Near infrared spectroscopy (NIRS) has been used to monitor oxygenation changes in muscle. Quantitative values for O₂ consumption, blood flow and venous saturation have been reported by several investigators. The amount of these measurements is, however, still limited and complete validation has not yet been established. The aim of this study was to investigate the different NIRS methods to calculate O₂ consumption (VO₂) and forearm blood flow (FBF) and to validate the data with the accepted method of strain-gauge plethysmography and blood sampling. Thirteen subjects were tested in rest and during static isometric handgrip exercise at 10% MVC. The NIRS optodes were positioned on the flexor region of the arm. A significant correlation was found between plethysmograph data and NIRS [tHb] during venous occlusion in rest (r EQ 0.925 - 0.994, P < 0.05) as well as during exercise (r equals 0.895 - 0.990, P < 0.05). No correlation was found, however, for the calculated FBF and VO₂ values between NIRS and the combination of plethysmography and blood sampling. In rest nor during exercise. It seems that although NIRS is a good qualitative monitoring technique, quantification is difficult due to the great variability that is found between the subjects.