Key result
Central arterial pressures estimated from the radial artery using the Pressure Recording Analytical Method (PRAM) correlated highly with directly measured aortic pressures in stable patients.
Why the study?
Central arterial pressure can be derived from peripheral waveforms, prompting comparison of central aortic pressures with radial pressures and PRAM-estimated central pressures.
Does the pressure recording analytical method (PRAM) accurately estimate central arterial pressure from the radial artery compared to direct intra-aortic measurement in patients undergoing invasive neuroradiological procedures?
Observational (n=21)
Single-blind
No
Does the pressure recording analytical method (PRAM) accurately estimate central arterial pressure from the radial artery compared to direct intra-aortic measurement in patients undergoing invasive neuroradiological procedures?
Answers are grounded in this paper and cite their sources.
PRAM can provide reliable estimates of central arterial pressure from radial artery waveforms in stable patients undergoing cerebral angiography, showing excellent correlation with direct aortic measurements.
Central arterial pressure can be derived from analysis of the peripheral artery waveform. The aim of this study was to compare central arterial pressures measured from an intra-aortic catheter with peripheral radial arterial pressures and with central arterial pressures estimated from the peripheral pressure wave using a pressure recording analytical method (PRAM). We studied 21 patients undergoing digital subtraction cerebral angiography under local or general anesthesia and equipped with a radial arterial catheter. A second catheter was placed in the ascending aorta for central pressure wave acquisition. Central (AO) and peripheral (RA) arterial waveforms were recorded simultaneously by PRAM for 90–180 s. During an off-line analysis, AO pressures were reconstructed (AOrec) from the RA trace using a mathematical model obtained by multi-linear regression analysis. The AOrec values obtained by PRAM were compared with the true central pressure value obtained from the catheter placed in the ascending aorta. Systolic, diastolic and mean pressures ranged from 79 to 180 mmHg, 47 to 102 mmHg, and 58 to 128 mmHg, respectively, for AO, and 83 to 174 mmHg, 47 to 107 mmHg, and 60 to 129 mmHg, respectively, for RA. The correlation coefficients between AO and RA were 0.86 (p < 0.01), 0.83 (p < 0.01) and 0.86 (p < 0.01) for systolic, diastolic and mean pressures, respectively, and the mean differences − 0.3 mmHg, 2.4 mmHg and 1.5 mmHg. The correlation coefficients between AO and AOrec were 0.92 (p < 0.001), 0.87 (p < 0.001) and 0.92 (p < 0.001), for systolic, diastolic and mean pressures, respectively, and the mean differences 0.01 mmHg, 1.8 mmHg and 1.2 mmHg. PRAM can provide reliable estimates of central arterial pressure.
No takes yet. Share an insight, caveat, or question.
Scolletta et al. (2019) conducted an observational in Critically ill neurological patients requiring digital subtraction cerebral angiography (n=21). Pressure Recording Analytical Method (PRAM) vs. Direct intra-aortic catheter measurement was evaluated on Correlation between measured and reconstructed systolic arterial pressure (r = 0.92, p=<0.001). Central arterial pressures estimated from the radial artery using the Pressure Recording Analytical Method (PRAM) correlated highly with directly measured aortic pressures in stable patients.
Synapse has enriched 10 related papers. Consider them for comparative context: