Raising the lower rate limit above the junctional rhythm's cycle length can restore biventricular pacing in patients experiencing loss of capture due to junctional rhythm.
Answers are grounded in this paper and cite their sources.
May restore biventricular pacing via simple rate adjustment in CRT patients with junctional rhythm; hypothesis-generating and requires prospective confirmation.
Disclaimer: Supplementary materials have been peer-reviewed but not copyedited. Figure S1. Rhythm during biventricular pacing. Figure S2. Restoration of biventricular pacing. Raising the lower rate limit to just slightly above the junctional rhythm's cycle length (793 ms vs 820 ms) restored biventricular pacing. For final device programming, a safety margin was added (85 beats/min, 705 ms) to ensure consistent biventricular pacing. Please note: The publisher is not responsible for the content or functionality of any supporting information supplied by the authors. Any queries (other than missing content) should be directed to the corresponding author for the article.