Why the study?
Atrial linear lesions are generally created with radiofrequency energy, leading researchers to evaluate the feasibility of cryothermal atrial linear ablation.
Does cryothermal atrial linear ablation improve conduction block and freedom from arrhythmias compared to radiofrequency ablation in patients with atrial fibrillation?
Does cryothermal atrial linear ablation improve conduction block and freedom from arrhythmias compared to radiofrequency ablation in patients with atrial fibrillation?
Cryoablation is effective for creating LA roof and CTI linear block but challenging for MI block, though it may result in fewer recurrent atrial tachycardias compared to radiofrequency ablation.
Cryoablation was associated with higher AF/AT freedom than radiofrequency; hypothesis-generating and requires randomized confirmation before practice change.
BACKGROUND: Atrial linear lesions are generally created with radiofrequency energy. We sought to evaluate the feasibility of cryothermal atrial linear ablation. METHODS AND RESULTS: Twenty-one atrial fibrillation (AF) patients underwent linear ablation on the left atrial (LA) roof, mitral isthmus (MI), and cavotricuspid isthmus (CTI) with 8-mm-tip cryocatheters following pulmonary vein isolation. The data were compared with those of 31 patients undergoing linear ablation with irrigated-tip radiofrequency catheters. Conduction block was successfully created in 18 of 20 (90%), 9 of 21 (43%), and 20 of 20 (100%) on the LA roof, MI, and CTI by endocardial cryoablation alone with 19.0 (12.0-24.0), 30.0 (23.0-34.0), and 14.0 (14.0-16.0) minute cryo applications, respectively. The presence of either an interposed circumflex artery or pouch at the MI was significantly associated with failed MI block (P = .04). Conduction block was created in 25 of 31 (83.9%), 27 of 31 (87.1%), and 30 of 31 (96.8%) on the roof, MI, and CTI, respectively, by radiofrequency ablation. During the 17.5 (13.0-31.7) months of follow-up, freedom from AF/atrial tachycardia (AT) was significantly higher in the cryo group (P = .05); especially, recurrent AT was more frequent in the RF group (8/31 vs 1/21; P = .03). Conduction block across the roof, MI, and CTI was durable in 6 of 12 (50.0%), 4 of 12 (33.3%), and 9 of 12 (75.0%) patients during second procedures. All nine patients (except one) with recurrent ATs had at least one roof or MI conduction resumption. CONCLUSIONS: Cryoablation is effective for creating a roof and CTI linear block, however, creating MI block by endocardial ablation alone was often challenging. Conduction resumption of LA linear block is common and recurrent arrhythmias, especially iatrogenic ATs, are more frequently observed after radiofrequency linear ablation.
No takes yet. Share an insight, caveat, or question.
Mukai et al. (2020) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: