Key result
Perosphere POC coagulometer detects therapeutic DOAC levels with >0.98 correlation to manual clotting time.
Why the study?
Readily available and rapid turn-around bedside assays to measure the anticoagulant effect of direct oral anticoagulants are lacking.
Does the Perosphere Technologies' PoC coagulometer accurately and precisely measure the anticoagulant effects of DOACs and enoxaparin compared to manual WBCT in spiked whole blood?
Does the Perosphere Technologies' PoC coagulometer accurately and precisely measure the anticoagulant effects of DOACs and enoxaparin compared to manual WBCT in spiked whole blood?
Effect estimate: R > 0.98
The Perosphere Technologies PoC coagulometer provides a rapid, sensitive, and precise bedside measurement of DOAC and enoxaparin anticoagulant effects in vitro, correlating strongly with manual whole blood clotting time.
Demonstrates strong in vitro correlation for Perosphere coagulometer; leaves open clinical applicability in DOAC-treated patients.
Readily available and rapid turn-around, bedside assays to measure the effect of the direct oral anticoagulants (DOACs) are not available. This study evaluates a point-of-care (PoC) coagulometer to assess the anticoagulant effects of the DOACs and low molecular weight heparin. Studies were done in fresh spiked blood from healthy volunteers. PoC coagulometer baseline clotting times were half that of the manual whole blood clotting time (WBCT, legacy method) and exhibited a sensitivity to DOAC anticoagulation approximately twice that of WBCT. All %CV values for both methods were < 10% with most being < 5% indicating acceptable precision of both methods. R2 values for both clotting time and percent rise from baseline were > 0.98 indicating a strong correlation between the two methods. Replicate measurements for all subjects showed a maximum upper %CV value of 5.56% and a maximum upper value of all absolute %Differences of 5.5%, with both criteria meeting predefined acceptance criteria. The dose–response curves for all subjects were linear across the concentration ranges tested. The Perosphere Technology PoC coagulometer detects a range of therapeutic levels of the DOACs apixaban, rivaroxaban, edoxaban, and dabigatran, as well as the low molecular weight heparin, enoxaparin, with high precision and sensitivity. Please check and confirm the processed affiliations are correct . and amend if necessary.
No takes yet. Share an insight, caveat, or question.
Bakhru et al. (2025) studied Healthy volunteers (in vitro blood spiking) (n=8). Point-of-care (PoC) coagulometer vs. Manual whole blood clotting time (WBCT) was evaluated on Correlation of clotting time and percent rise of clotting time from baseline between methods (R > 0.98). The Perosphere point-of-care coagulometer detected therapeutic levels of direct oral anticoagulants with high precision and strong correlation (R > 0.98) to manual whole blood clotting time.