Key result
Novel calibration-free electrochemical sensor quantifies rivaroxaban and edoxaban in under 30 seconds.
Why the study?
Quantifying DOACs below 30 ng mL−1 is crucial in emergency care centres when patients require emergency surgery or reversal due to bleeding or ischemic stroke.
Can a calibration-free electrochemical sensor accurately quantify factor-Xa inhibitors in blood plasma at the point-of-care?
Can a calibration-free electrochemical sensor accurately quantify factor-Xa inhibitors in blood plasma at the point-of-care?
A novel calibration-free electrochemical sensor can rapidly quantify DOACs and LMWHs at clinically relevant low concentrations, offering potential for point-of-care testing in emergencies.
May enable rapid point-of-care DOAC testing; leaves open translation to human samples and clinical use.
This article presents a novel proof of concept for the blood plasma quantification of clinically relevant concentrations of direct oral anticoagulants, DOACs, including rivaroxaban and edoxaban, as well as low-molecular-weight heparins, LMWHs, such as enoxaparin and dalteparin, utilising a calibration-free disposable electrochemical sensor with co-facing electrodes. A dose-response curve was generated for rivaroxaban and edoxaban to demonstrate the sensor's ability to detect ≥9.00 ng mL−1 rivaroxaban and quantify it in the 11.0–140 ng mL−1 range. Similarly, the lower detection limit for edoxaban was 12.9 ng mL−1, with a quantification range of 16.8–140 ng mL−1. The significance of this sensor lies in its ability to quantify rivaroxaban and edoxaban below 30 ng mL−1, which is crucial in emergency care centres when patients undergoing DOAC therapy require emergency surgery or reversal of DOACs due to bleeding or ischemic stroke. Furthermore, the sensor can detect ≥0.016 IU mL−1 enoxaparin and ≥0.013 IU mL−1 dalteparin and quantify them in the 0.025–0.75 and 0.019–0.75 IU mL−1 range, respectively. Additionally, a dose-response curve was presented to demonstrate the potential ability of this sensor to quantify factor-Xa inhibitors independently of which DOACs or LMWHs are used. With the assay completed in less than 30 s using a minimal volume of 7 μL sample, the possibility to work at physiological pH and under calibration-free format makes this assay an excellent candidate for point-of-care testing.
No takes yet. Share an insight, caveat, or question.
Mruthunjaya et al. (2023) studied Anticoagulation therapy monitoring. Calibration-free disposable electrochemical sensor was evaluated on Quantification of DOACs and LMWHs. A novel calibration-free electrochemical sensor successfully detected and quantified rivaroxaban (≥9.00 ng/mL) and edoxaban (≥12.9 ng/mL) in under 30 seconds using a 7 μL sample.