The relationship between systolic blood pressure and pulse transit time in anesthetized dogs was nonlinear and logarithmic, with correlation coefficients often greater than 0.96.
The relationship between systolic blood pressure and pulse transit time is nonlinear and logarithmic, suggesting caution is necessary when using this relationship to monitor blood pressure due to variability in coefficients.
Answers are grounded in this paper and cite their sources.
Nonlinear SBP-PTT relationship in dogs urges caution for monitoring applications; leaves open clinical translation and validation.
Pulse wave transit time (PTT), as measured between the ECG R-wave and the arterial pulse wave at a distal site, has been previously reported to linearly correlate with blood pressure (BP). This study reinvestigated the relationship where various doses of epinephrine were administered to three anesthetized dogs to invoke a wide range of hypotension. The invasive systolic blood pressure in the left ventricle, the noninvasive photoplethysmogram pulse wave and the ECG of lead II were recorded synchronously. The results indicated that the relationship between the systolic BP and the PTT was nonlinear and logarithmic: P=b 1 -b 2 lnT, where b 1 and b 2 are coefficients. The estimated BP using the nonlinear relationship was highly correlated with the measured BP; the correlation coefficients were often greater than 0.96 in the 36 trials. The root mean square of the estimation error was less than 6.53 mmHg. However, the coefficients b 1 and b 2 varied between trials, doses, and dogs. An analysis of the error contribution revealed that the estimation error increased linearly with the coefficient errors, Δb 1 and Δb 2 , and with the PTT error, ΔT. The definition of the pulse arrival time (PAT) also affected the performance of the estimation. The performance based on five PAT definitions was compared. This study shows that caution is necessary when using this relationship to monitor BP.
No takes yet. Share an insight, caveat, or question.
Tang et al. (2014) studied Blood pressure monitoring (n=3). Pulse transit time (PTT) vs. Invasive systolic blood pressure was evaluated on Relationship between systolic blood pressure and pulse transit time. The relationship between systolic blood pressure and pulse transit time in anesthetized dogs was nonlinear and logarithmic, with correlation coefficients often greater than 0.96.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: