Key result
PPG-PTT outperforms PAT for tracking systolic blood pressure, achieving an R2 of ~0.85.
Why the study?
Does PPG-PTT provide a more accurate noninvasive estimate of blood pressure compared to PAT?
Does PPG-PTT provide a more accurate noninvasive estimate of blood pressure compared to PAT?
Answers are grounded in this paper and cite their sources.
PPG-PTT tracks acute blood pressure changes significantly better than the more commonly used PAT, offering a promising avenue for cuff-less BP monitoring.
Pulse transit time (PTT) measured as the time delay between invasive proximal and distal blood pressure (BP) or flow waveforms (invasive PTT [I-PTT]) tightly correlates with BP PTT estimated as the time delay between noninvasive proximal and distal arterial waveforms could therefore permit cuff-less BP monitoring. A popular noninvasive PTT estimate for this application is the time delay between ECG and photoplethysmography (PPG) waveforms (pulse arrival time [PAT]). Another estimate is the time delay between proximal and distal PPG waveforms (PPG-PTT). PAT and PPG-PTT were assessed as markers of BP over a wide physiologic range using I-PTT as a reference. Waveforms for determining I-PTT, PAT, and PPG-PTT through central arteries were measured from swine during baseline conditions and infusions of various hemodynamic drugs. Diastolic, mean, and systolic BP varied widely in each subject (group average (mean ± SE) standard deviation between 25 ± 2 and 36 ± 2 mmHg). I-PTT correlated well with all BP levels (group average R(2) values between 0.86 ± 0.03 and 0.91 ± 0.03). PPG-PTT also correlated well with all BP levels (group average R(2) values between 0.81 ± 0.03 and 0.85 ± 0.02), and its R(2) values were not significantly different from those of I-PTT PAT correlated best with systolic BP (group average R(2) value of 0.70 ± 0.04), but its R(2) values for all BP levels were significantly lower than those of I-PTT (P < 0.005) and PPG-PTT (P < 0.02). The pre-ejection period component of PAT was responsible for its inferior correlation with BP In sum, PPG-PTT was not different from I-PTT and superior to the popular PAT as a marker of BP.
No takes yet. Share an insight, caveat, or question.
Gao et al. (2016) studied Healthy (swine model) (n=6). Photoplethysmography pulse transit time (PPG-PTT) vs. Pulse arrival time (PAT) and invasive PTT was evaluated on Correlation (group average R2) with systolic blood pressure (p=<0.02). PPG-PTT was superior to PAT as a marker of systolic blood pressure and was not different from invasive PTT.
Synapse has enriched 10 related papers. Consider them for comparative context: