Why the study?
Does pulse arrival time (PAT) accurately track diastolic pressure compared to pulse transit time (PTT) in a canine model?
Does pulse arrival time (PAT) accurately track diastolic pressure compared to pulse transit time (PTT) in a canine model?
Pulse transit time is superior to pulse arrival time for tracking diastolic pressure in a canine model.
Answers are grounded in this paper and cite their sources.
PTT may outperform PAT for diastolic pressure tracking in canines; leaves open human validation before clinical use.
We compared pulse arrival time (PAT) and pulse transit time (PTT) in terms of their ability to track diastolic pressure (DP). We performed the comparison using high fidelity, invasive arterial waveforms recorded from six dogs during multiple interventions. On average, DP ranged from 40 to 106 mmHg and therefore varied widely. PAT and PTT were able to predict DP with average root-mean-squared-errors of 9.8 ± 5.8 mmHg and 5.7 ± 2.0 mmHg (p = 0.02). Thus, even though PAT is simpler to measure, we can only recommend using PTT for tracking DP.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2011) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: