Key result
The MediGuide catheter tracking system significantly reduced fluoroscopy time for double transseptal puncture compared to the conventional approach (0.48 vs. 5.9 minutes; P<0.0001).
Why the study?
Does a nonfluoroscopic catheter tracking system reduce fluoroscopy time during transseptal puncture in patients undergoing radiofrequency ablation for atrial fibrillation?
Observational (n=32)
Does a nonfluoroscopic catheter tracking system reduce fluoroscopy time during transseptal puncture in patients undergoing radiofrequency ablation for atrial fibrillation?
Absolute Event Rate: 0.48% vs 5.9%
p-value: p=<0.0001
A nonfluoroscopic catheter tracking system significantly reduces fluoroscopy time during transseptal puncture for atrial fibrillation ablation without increasing complications.
Supports reduced fluoroscopy in AF ablation; leaves open randomized confirmation before practice change.
BACKGROUND: Radiation exposure in the electrophysiology lab is a major occupational hazard to the electrophysiologists. A catheter localization system (MediGuide Technology, St. Jude Medical Inc., St. Paul, MN, USA) allows the integration of electroanatomical mapping and x-ray imaging, and has been shown to be effective in reducing radiation exposure during several electrophysiological procedures. We intended to evaluate the feasibility of this catheter tracking system to guide transseptal (TS) access. METHODS: The feasibility of performing TS puncture with MediGuide (MDG) was assessed in a prospective observational study in 16 patients undergoing radiofrequency ablation for atrial fibrillation. These patients were compared to 16 matched patients undergoing similar procedures during the same time frame using conventional approach. There were no differences in mean age, gender distribution, and body mass index between the two groups. Total duration of fluoroscopic exposure during TS puncture was compared between the two groups. RESULTS: All patients underwent successful TS puncture. Fluoroscopy time for double TS puncture using the MDG system was significantly lower than the control group (0.48 ± 0.17 minutes vs. 5.9 ± 0.65 minutes; P < 0.0001). No major complications occurred during the procedures in either group. CONCLUSIONS: TS puncture can be successfully performed using MDG, and results in significant reduction in radiation exposure.
No takes yet. Share an insight, caveat, or question.
Mansour et al. (2015) conducted an observational in Atrial fibrillation (n=32). MediGuide catheter tracking system vs. Conventional approach was evaluated on Fluoroscopy time for double transseptal puncture (p=<0.0001). The MediGuide catheter tracking system significantly reduced fluoroscopy time for double transseptal puncture compared to the conventional approach (0.48 vs. 5.9 minutes; P<0.0001).