Key result
Ultrasound elasticity measurements of the posterior wall within the intima-media complex visible range differed by up to 6% in a phantom experiment.
Why the study?
In carotid artery ultrasound elasticity measurement, the intima-media complex is visible near as well as along the central axis of the short-axis view, raising questions about measurement accuracy across this visible range.
Does measurement position within the intima-media complex visible range affect the accuracy of ultrasound elasticity measurement in a phantom model?
Does measurement position within the intima-media complex visible range affect the accuracy of ultrasound elasticity measurement in a phantom model?
For the highest accuracy in ultrasound elasticity measurement of the carotid artery, it is important to measure along the central axis of the short-axis view.
Measurement position may influence elasticity accuracy; leaves open standardization of carotid ultrasound protocols.
Studies to investigate the ultrasound elasticity measurement of the carotid artery, for early detection of arteriosclerosis, are ongoing. In the long-axis cross-sectional measurement in vivo, the position where the intima-media complex (IMC) is visible on the B-mode image was assumed to be the central axis of the short-axis view of the carotid artery. However, the IMC is also visible near the central axis of the short-axis view of the carotid artery. In the present study, accuracy in elasticity measurement within the IMC visible range was evaluated through a phantom experiment. The elasticities of the posterior wall measured at plural points within the IMC visible range differed by up to 6%. From the experimental results, we concluded that for the highest accuracy, it is important to measure along the central axis of the short-axis view of the carotid artery.
No takes yet. Share an insight, caveat, or question.
Akiyama et al. (2021) studied arteriosclerosis. ultrasound elasticity measurement was evaluated on accuracy in elasticity measurement. Ultrasound elasticity measurements of the posterior wall within the intima-media complex visible range differed by up to 6% in a phantom experiment.