Key result
Miniature epicardial ultrasound transducers detected regional myocardial ischaemia during LAD occlusion, indicated by a decrease in systolic velocity from 0.9 to 0.1 cm/s (P<0.01).
Why the study?
Can miniature epicardial ultrasound transducers detect regional myocardial ischemia in a porcine model?
Can miniature epicardial ultrasound transducers detect regional myocardial ischemia in a porcine model?
p-value: p=<0.01
Miniature epicardial ultrasound transducers provide continuous, sensitive detection of regional myocardial ischemia in a porcine model, suggesting potential utility for monitoring during cardiac surgery.
Feasible for ischemia detection in porcine model; leaves open translation to intraoperative human monitoring.
BACKGROUND: Detection of myocardial ischaemia during and after cardiac surgery remains a challenge. Echocardiography is more sensitive in ischaemia detection than echocardiography (ECG) and haemodynamic monitoring, but demands repeated examinations for monitoring over time. We have developed and validated an ultrasonic system that permits continuous real-time assessment of myocardial ischaemia using miniature epicardial ultrasound transducers. METHODS: In an open-chest porcine model (n=8), prototype ultrasound transducers were fixed on the epicardium in the left anterior descending and circumflex coronary artery supply regions, providing continuous measurement of transmural myocardial velocities. Peak systolic velocity and post-systolic velocity were recorded simultaneously with ECG, left ventricular pressure and arterial pressure. Two-dimensional (2D) echocardiographic strain was used as a reference. Global changes were induced by infusing fluid, epinephrine, nitroprusside and esmolol. Regional changes were induced by occluding the left anterior descending coronary artery (LAD). Subsequent LAD stenosis was performed in a subgroup, with flow reduction to 50% of baseline level and further to occlusion. RESULTS: Systolic velocity in the LAD region decreased during LAD occlusion (0.9+/-0.1 to 0.1+/-0.1 cm s(-1), P<0.01), whereas post-systolic velocity increased (0.3+/-0.1 to 2.3+/-0.1 cm s(-1), P<0.01). No changes occurred in the circumflex coronary artery (CX) region. Severe ischaemia was confirmed by reduction in 2D echocardiography strain calculations. Changes in myocardial velocities assessed by miniature transducer during ischaemia differed from changes during all global interventions. Significant reduction in systolic velocity occurred at 50% LAD flow (0.9+/-0.1 to 0.5+/-0.1 cm s(-1), P=0.02) with further decrease on following occlusion (0.0+/-0.0 cm s(-1), P<0.01). Post-systolic velocity increased both from baseline to 50% LAD flow, and further to occlusion. CONCLUSION: The epicardial transducers provided continuous assessment of regional myocardial function and detected ischaemia with high sensitivity and specificity. Further development of this system may provide a useful tool for myocardial monitoring during and after cardiac surgery.
No takes yet. Share an insight, caveat, or question.
Espinoza et al. (2009) studied Myocardial ischaemia (n=8). Miniature epicardial ultrasound transducers vs. Baseline was evaluated on Systolic and post-systolic velocity (p=<0.01). Miniature epicardial ultrasound transducers detected regional myocardial ischaemia during LAD occlusion, indicated by a decrease in systolic velocity from 0.9 to 0.1 cm/s (P<0.01).