Key result
Omron HeartGuide smartwatch closely tracks conventional pulse rate monitoring within ~1.5 bpm.
Why the study?
While oscillometric blood pressure measurement is established, its use for pulse rate monitoring via wrist-worn devices remains largely unexplored.
Does the Omron HeartGuide smartwatch accurately measure pulse rate compared to a conventional upper-arm ambulatory blood pressure monitor in adult outpatients?
Observational (n=50)
Open-label
Yes
Does the Omron HeartGuide smartwatch accurately measure pulse rate compared to a conventional upper-arm ambulatory blood pressure monitor in adult outpatients?
Mean Difference: 1.5
Wrist-worn oscillometric smartwatches can measure pulse rate with high accuracy compared to upper-arm devices, supporting their potential use in out-of-hospital telemetric monitoring, though algorithmic adjustments are needed for patients with arrhythmias.
Supports outpatient pulse monitoring with wrist devices; leaves open prospective validation in larger, arrhythmia-enriched cohorts.
BACKGROUND: Monitoring pulse rate is fundamental to cardiovascular health management and early detection of rhythm disturbances. While oscillometric blood pressure measurement is well established and validated in clinical practice, its use for pulse rate monitoring, particularly via wrist-worn devices, remains largely unexplored. OBJECTIVE: This study investigates whether a smartwatch that performs oscillometric blood pressure measurements at the wrist can also deliver reliable pulse rate readings using the same method. METHODS: This study compared pulse rates recorded by the Omron HeartGuide smartwatch and conventional ambulatory blood pressure monitors in 50 patients over 24 h. Measurements were taken consecutively, and data were analyzed using intraclass correlation coefficients (ICCs) and Bland-Altman plots. RESULTS: The study showed a high ICC of 0.971, indicating excellent agreement between devices. The average pulse rate difference was 1.5 bpm, with the Omron HeartGuide reporting slightly lower rates, especially in patients with atrial fibrillation. CONCLUSIONS: This study demonstrates that oscillometric pulse-rate monitoring at the wrist can achieve a high degree of accuracy, comparable to conventional upper-arm devices. Given that oscillometric smartwatches like the Omron HeartGuide are already used for blood pressure monitoring, the findings suggest that they may also be suitable for pulse rate measurement, potentially enhancing their role in telemetric healthcare, but further research is needed, particularly in patients with arrhythmias.
No takes yet. Share an insight, caveat, or question.
Vaseekaran et al. (2025) conducted an observational in Patients scheduled for 24-hour ambulatory blood pressure monitoring (n=50). Omron HeartGuide smartwatch vs. Conventional ambulatory blood pressure monitor (Spacelabs 90217A-2) was evaluated on Pulse rate agreement (Mean difference 1.5 bpm (ICC 0.971)). The Omron HeartGuide smartwatch showed excellent agreement for pulse rate measurement with an intraclass correlation coefficient of 0.971 and a mean difference of 1.5 bpm compared to a conventional monitor.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: