Key result
Omron HeartGuide underestimates resting systolic BP by ~10 mmHg compared to Welch Allyn ProBP.
Why the study?
A new watch-type device, the Omron HeartGuide, claims to provide clinically accurate blood pressure measurement while tracking activity and sleep, but its performance required evaluation.
Does the Omron HeartGuide watch-type blood pressure monitor provide accurate blood pressure measurements compared to the Welch Allyn ProBP 3400?
Observational
Does the Omron HeartGuide watch-type blood pressure monitor provide accurate blood pressure measurements compared to the Welch Allyn ProBP 3400?
The Omron HeartGuide watch-type blood pressure monitor significantly underestimates blood pressure compared to a validated clinical monitor and is not a suitable replacement for existing devices.
Answers are grounded in this paper and cite their sources.
Clinicians should not substitute the Omron HeartGuide for validated monitors; this Level 5 comparison leaves open its accuracy in broader use.
Hypertension affects approximately 100 million U.S. adults and is the leading single contributing risk factor to all-cause mortality. Accurate blood pressure (BP) measurement is essential in the treatment of BP, and a number of devices exist for monitoring. Recently, a new watch-type design was released, the Omron HeartGuide (BP8000), with claims to provide clinically accurate BP measurement while also tracking activity and sleep similar to smart watches. The aim of this research was done in two studies: (1) evaluation of the HeartGuide device for measurement of resting BP and heart rate (HR); and (2) assessment of the HeartGuide for BP, HR, step-counting and sleep monitoring during activities of daily living. Study 1 compared the Omron HeartGuide to the previously validated Welch Allyn ProBP 3400 following a modified version of the Universal Standard for validation of BP measuring devices set by the AAMI/ESH/ISO. While resting HR measured by the HeartGuide was similar to Welch Allyn measures, both systolic and diastolic BP were significantly lower ( P ≤0.001), with differences of 10.4 (11.1) and 3.2 (10.0) mmHg, respectively. Study 2 compared HeartGuide measures to Welch Allyn measures for BP, HR, steps and sleep during various body positions (supine, seated, standing), physiological stressors (cold pressor test, lower body submersion, exercise), and free-living. The HeartGuide significantly underestimated BP though provided accurate HR during most conditions. It also significantly underestimated steps, but reported sleep measures similar to those subjectively reported. Based on the significant differences between the HeartGuide and Welch Allyn, our data indicate the HeartGuide is not a suitable replacement for existing BP monitors.
No takes yet. Share an insight, caveat, or question.
Harfmann et al. (2023) reported an observational. Omron HeartGuide (BP8000) vs. Welch Allyn ProBP 3400 was evaluated on Resting systolic and diastolic blood pressure (Mean difference -10.4 mmHg (systolic) and -3.2 mmHg (diastolic), p=≤0.001). The Omron HeartGuide underestimated resting systolic and diastolic blood pressure compared to the Welch Allyn ProBP 3400 by 10.4 mmHg and 3.2 mmHg, respectively.