Key result
Cardiac resynchronization therapy upgrade improved left ventricular apex contraction and resolved systolic rebound stretch.
Why the study?
Does a CRT-P upgrade improve ventricular hypokinesis and dyssynchrony induced by apical pacing in a patient with complex congenital heart disease?
Case Report (n=1)
Does a CRT-P upgrade improve ventricular hypokinesis and dyssynchrony induced by apical pacing in a patient with complex congenital heart disease?
CRT upgrade can successfully treat pacing-induced intraventricular dyssynchrony and ventricular hypokinesis at the apical pacing site in patients with complex congenital heart disease.
Answers are grounded in this paper and cite their sources.
May support CRT-P upgrade for apical pacing dyssynchrony in complex CHD; hypothesis-generating and requires prospective validation.
The left ventricular (LV) apex is recommended as the first choice for positioning the epicardial pacing. We encountered a patient with congenital heart disease (CHD) showing hypokinesis of the LV apical pacing site after implantation of a pacemaker with epicardial leads. This phenomenon was revealed by the early shortening and systolic rebound stretch of the same lesion on two-dimensional speckle tracking echocardiography, which developed in the intraventricular dyssynchrony between the LV apex and base. Cardiac resynchronization therapy provided an excellent result around the hypokinetic lesion. It is wise to arrange detailed evaluations in each patient with complicated CHD, aiming at a successful treatment to enable ventricular synchronicity.
No takes yet. Share an insight, caveat, or question.
Suzuki et al. (2022) conducted a case report in Congenital heart disease with Fontan physiology and pacing-induced ventricular hypokinesis (n=1). Cardiac resynchronization therapy (CRT-P upgrade) vs. Conventional pacemaker with LV apical pacing (baseline) was evaluated on Left ventricular apex contraction (fractional area change) and intraventricular dyssynchrony. Cardiac resynchronization therapy upgrade improved left ventricular apex contraction and resolved systolic rebound stretch.
Synapse has enriched 10 related papers. Consider them for comparative context: