Key result
Noninvasive arterial pressure measurement using ClearSight during anesthesia induction for cardiac surgery showed a mean difference of -5.47 mmHg for mean arterial pressure compared to invasive measurement.
Why the study?
The study compared noninvasive continuous arterial pressure monitoring using the ClearSight vascular-unloading technique with invasive arterial pressure measurement during anesthesia induction for major cardiac surgery.
Does noninvasive continuous arterial pressure monitoring using ClearSight accurately reflect invasive arterial pressure measurements during anesthesia induction in patients undergoing major cardiac surgery?
Population
54 patients undergoing major cardiac surgery
Comparison
ClearSight noninvasive arterial pressure measurement vs invasive arterial pressure measurement
Design
Prospective monocentric study
Follow-up
Mean time period of 27 minutes
Authors
Loading...
Noninvasive systolic bias warrants caution on induction; leaves open validation before replacing invasive monitoring in cardiac surgery.
Observational (n=54)
No
Does noninvasive continuous arterial pressure monitoring using ClearSight accurately reflect invasive arterial pressure measurements during anesthesia induction in patients undergoing major cardiac surgery?
Mean Difference: -5.47 (95% CI -27.2–16.26)
p-value: p=<0.05
Noninvasive continuous arterial pressure monitoring using ClearSight provides acceptable accuracy for mean arterial pressure compared to invasive monitoring during anesthesia induction in major cardiac surgery.
Frank et al. (2021) conducted an observational in Major cardiac surgery (n=54). ClearSight noninvasive arterial pressure measurement vs. Invasive arterial pressure measurement was evaluated on Mean difference in mean arterial pressure (MAP) (MD -5.47 mmHg, 95% CI -27.20 to 16.26, p=<0.05). Noninvasive arterial pressure measurement using ClearSight during anesthesia induction for cardiac surgery showed a mean difference of -5.47 mmHg for mean arterial pressure compared to invasive measurement.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: