Key result
Noninvasive ClearSight System shows acceptable agreement versus invasive radial pressure with a ~4 mmHg bias.
Why the study?
The validity of noninvasive continuous arterial pressure measurement using the ClearSight System during induction of anesthesia in patients undergoing cardiovascular surgery had not been evaluated.
Does noninvasive continuous arterial pressure measurement by the ClearSight System accurately reflect invasive radial arterial pressure during induction of anesthesia for cardiovascular surgery?
Observational (n=18)
No
Does noninvasive continuous arterial pressure measurement by the ClearSight System accurately reflect invasive radial arterial pressure during induction of anesthesia for cardiovascular surgery?
The ClearSight System provides acceptable accuracy for mean arterial pressure, but not systolic or diastolic pressure, compared to invasive radial arterial monitoring during anesthesia induction for cardiovascular surgery.
Answers are grounded in this paper and cite their sources.
ClearSight may suffice for mean arterial pressure but not systolic or diastolic during induction; leaves open prospective validation versus invasive monitoring in cardiovascular surgery.
BACKGROUND: Since blood pressure tends to be unstable during induction of anesthesia in patients undergoing cardiovascular surgery, an artery catheter is often inserted before induction to continuously monitor arterial pressure during induction of anesthesia. ClearSight System™ enables noninvasive continuous measurement of beat-to-beat arterial pressure via a single finger cuff without pain using photoplethysmographic technology. If ClearSight System™ can replace intra-arterial pressure measurement, blood pressure could be easily and noninvasively assessed. However, the validity of ClearSight System™ during induction of anesthesia in patients undergoing cardiovascular surgery has not been evaluated. The aim of this study was to compare blood pressure measured by ClearSight System™ with intra-arterial pressure during induction of anesthesia for cardiovascular surgery. METHODS: This study was registered retrospectively. Data during induction of anesthesia for elective cardiovascular surgery were obtained for patients in whom noninvasive arterial pressure was measured by ClearSight System™ (APcs) and invasive radial arterial pressure (APrad) was measured simultaneously. According to the widely used criteria formulated by international standards from the Association for the Advancement of Medical Instrumentation, the acceptable bias and precision for arterial pressure measurements were fixed at < 5 mmHg and 8 mmHg, respectively. RESULTS: Data for 18 patients were analyzed. For 3068 analyzed paired measurements, values of APcs vs APrad bias (precision) were 13.2 (17.5), - 9.1 (7.3) and - 3.9 (7.8) mmHg for systolic, diastolic, and mean arterial pressures, respectively. CONCLUSIONS: Mean arterial pressure measured by ClearSight System™ could be considered as an alternative for mean radial arterial pressure during induction of anesthesia for elective cardiovascular surgery.
No takes yet. Share an insight, caveat, or question.
Tanioku et al. (2020) conducted an observational in Elective cardiovascular surgery (n=18). ClearSight System noninvasive arterial pressure measurement vs. Invasive radial arterial pressure measurement was evaluated on Bias and precision between noninvasive and invasive mean arterial pressure (bias -3.9 mmHg (precision 7.8 mmHg)). Mean arterial pressure measured by the ClearSight System showed a bias of -3.9 mmHg and precision of 7.8 mmHg compared to invasive radial arterial pressure, meeting AAMI criteria for acceptable agreement.