Key result

Non-invasive blood pressure monitoring significantly overestimated systolic pressure and underestimated diastolic and mean arterial pressures compared to invasive monitoring during high-risk surgeries.

Why the study?

Intraoperative blood pressure monitoring guides hemodynamic management, but there is a need to compare non-invasive and invasive blood pressure monitoring methods in patients undergoing high-risk prolonged surgeries.

Does non-invasive blood pressure monitoring accurately reflect invasive blood pressure monitoring in patients undergoing high-risk prolonged surgeries?

Population

40 patients undergoing high-risk prolonged surgeries

Comparison

Non-invasive oscillometric blood pressure monitoring vs invasive blood pressure monitoring via radial artery line

Design

Comparative observational study

Authors

UDUtpal Kumar DuttaMBMahin Aboobacker BRBRituparna Bora

Discussion

Loading...

Member takes

Implication

NIBP differs significantly from invasive readings in high-risk surgery; leaves open whether routine invasive monitoring improves outcomes pending larger trials.

Study Design

Type

Observational (n=40)

Multicenter

No

Structured PICO

Does non-invasive blood pressure monitoring accurately reflect invasive blood pressure monitoring in patients undergoing high-risk prolonged surgeries?

P
Population
40 patients undergoing high-risk prolonged surgeries.
E
Exposure
Non-invasive blood pressure (NIBP) monitoring by oscillometric method
C
Comparator
Invasive blood pressure (IBP) monitoring via radial artery line
O
Outcome
Agreement between non-invasive and invasive blood pressure measurements (systolic, diastolic, and mean arterial pressure) assessed by Bland-Altman plotsurrogate

Invasive blood pressure monitoring remains recommended for high-risk surgeries due to significant discrepancies (overestimation of systolic and underestimation of diastolic/MAP) when using non-invasive oscillometric methods.

Limitations

  • Small sample size requiring a large-scale study to arrive at a definitive conclusion
  • Small sample size (large-scale study required to arrive at a definitive conclusion)

Cite This Study

Dutta et al. (2021) conducted an observational in High-risk prolonged surgeries (n=40). Non-invasive blood pressure monitoring vs. Invasive blood pressure monitoring was evaluated on Blood pressure measurement comparison (systolic, diastolic, and mean arterial pressure). Non-invasive blood pressure monitoring significantly overestimated systolic pressure and underestimated diastolic and mean arterial pressures compared to invasive monitoring during high-risk surgeries.

synapsesocial.com/papers/6a9cfcc9f4d73a8d78599ae9https://doi.org/10.18410/jebmh/2021/512
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Invasive and Concomitant Noninvasive Intraoperative Blood Pressure Monitoring2011 · 176 citations
  2. 2Recommendations for standards of monitoring during anaesthesia and recovery 2015 : Association of Anaesthetists of Great Britain and Ireland2015 · 490 citations
  3. 3Comparison of Continuous Noninvasive Blood Pressure Monitoring by TL-300 With Standard Invasive Blood Pressure Measurement in Patients Undergoing Elective Neurosurgery2015 · 19 citations
  4. 4Measuring blood pressure: pitfalls and recommendations2004 · 76 citations
  5. 5Catheter replacement of the needle in percutaneous arteriography2008 · 466 citations