Key result
Continuous noninvasive blood pressure monitoring by TL-300 demonstrated high accuracy compared to invasive measurement, with a mean BP bias of 1.8 mm Hg (95% limits of agreement -6.4 to +10.1 mm Hg).
Why the study?
Does continuous noninvasive blood pressure monitoring by TL-300 provide accurate and precise measurements compared to standard invasive blood pressure measurement in patients undergoing elective neurosurgery?
Population
23 patients undergoing elective neurosurgery
Comparison
Continuous noninvasive blood pressure monitoring… vs Standard invasive blood pressure measurement via…
Design
Cohort
Follow-up
intraoperative
Authors
Loading...
May support TL-300 for noninvasive BP monitoring in neurosurgery; leaves open need for randomized validation before practice change.
Observational (n=23)
Does continuous noninvasive blood pressure monitoring by TL-300 provide accurate and precise measurements compared to standard invasive blood pressure measurement in patients undergoing elective neurosurgery?
Mean Difference: 1.8 (95% CI -6.4–10.1)
The TL-300 system is a promising noninvasive alternative to invasive arterial catheters for intraoperative blood pressure monitoring, demonstrating high accuracy and precision.
Lin et al. (2015) conducted an observational in Elective neurosurgery (n=23). Continuous noninvasive blood pressure monitoring by TL-300 vs. Standard invasive radial arterial catheter method was evaluated on Accuracy and precision of continuous BP monitoring (mean BP) (Bias 1.8 mm Hg, 95% CI -6.4 to 10.1). Continuous noninvasive blood pressure monitoring by TL-300 demonstrated high accuracy compared to invasive measurement, with a mean BP bias of 1.8 mm Hg (95% limits of agreement -6.4 to +10.1 mm Hg).