Key result
Continuous noninvasive blood pressure monitoring by TL-300 demonstrated high accuracy compared to invasive measurement, with a mean BP bias of 1.8 mm Hg.
Why the study?
Does continuous noninvasive blood pressure monitoring by TL-300 provide accurate and precise measurements compared to standard invasive blood pressure measurement in patients undergoing elective neurosurgery?
Observational (n=23)
Does continuous noninvasive blood pressure monitoring by TL-300 provide accurate and precise measurements compared to standard invasive blood pressure measurement in patients undergoing elective neurosurgery?
The TL-300 system is a promising noninvasive alternative to invasive arterial catheters for intraoperative blood pressure monitoring, demonstrating high accuracy and precision.
Answers are grounded in this paper and cite their sources.
May support TL-300 for noninvasive BP monitoring in neurosurgery; leaves open need for randomized validation before practice change.
BACKGROUND: Intraoperative blood pressure (BP) is one of the basic vital signs monitoring. Compared with standard invasive BP measurement, TL-300 allows for a continuous and beat-to-beat noninvasive intraoperative BP monitoring. The current retrospective study compared the accuracy and precision of this noninvasive technique for continuous BP monitoring with that of standard invasive BP measurement in patients undergoing elective neurosurgery. MATERIALS AND METHODS: BP records of 23 patients undergoing elective neurosurgery, measured by both noninvasive TL-300 and invasive radial arterial catheter method, were retrospectively analyzed. Variability in BP data was analyzed by using linear regressions and Bland-Altman analysis. RESULTS: Four thousand three hundred eighty-one pairs of BP measurements from a total of 23 patients were included. The coefficient of determination of systolic, diastolic, and mean BP were 0.908, 0.803, and 0.922, respectively. And their bias was found to be 1.3±5.87 mm Hg (95% limits of agreement: -10.2 to +12.8 mm Hg), 2.8±6.40 mm Hg (95% limits of agreement: -9.8 to +15.3 mm Hg), and 1.8±4.20 mm Hg (95% limits of agreement: -6.4 to +10.1 mm Hg), respectively. CONCLUSIONS: TL-300 system is a promising noninvasive alternative to the invasive arterial catheter method for intraoperative BP monitoring, with a high accuracy and precision. With the limitation of the current retrospective study, further prospective method comparison studies are needed.
No takes yet. Share an insight, caveat, or question.
Lin et al. (2015) conducted an observational in Elective neurosurgery (n=23). Continuous noninvasive blood pressure monitoring by TL-300 vs. Standard invasive radial arterial catheter method was evaluated on Accuracy and precision of continuous BP monitoring (mean BP) (Bias 1.8 mm Hg, 95% CI -6.4 to 10.1). Continuous noninvasive blood pressure monitoring by TL-300 demonstrated high accuracy compared to invasive measurement, with a mean BP bias of 1.8 mm Hg.