Answers are grounded in this paper and cite their sources.
Two versions of a new high-resolution X-ray computed tomography system are being developed to screen mutagenized mice in the Oak Ridge National Laboratory Mammalian Genetics Research Facility. The first prototype employs a single-pixel CdZnTe detector with a pinhole collimator operating in pulse counting mode. The second version employs a phosphor screen/CCD detector operating in current mode. The major system hardware includes a low-energy X-ray tube, two linear translation stages and a rotational stage. For the single-pixel detector, image resolution is determined by the step size of the detector stage; preliminary images have been acquired at 100 /spl mu/m and 250 /spl mu/m resolutions. The resolution of the phosphor screen detector is determined by the modulation transfer function of the phosphor screen; images with resolutions approaching 50 /spl mu/m have been acquired. The system performance with the two detectors is described and recent images are presented.
No takes yet. Share an insight, caveat, or question.
Paulus et al. (1999) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: