A BC523A liquid scintillator loaded with boron-10 was tested as a detector for both fast and thermal neutrons. A pulse shape discrimination (PSD) method based on a zero-crossing principle was applied to distinguish between neutron and gamma radiation. High quantum efficiency Photonis XP5500B photomultiplier was used to enhance light detection from the scintillator. This allowed a good registration of the energy spectrum of neutron capture events on boron-10, corresponding to about 60 keV gamma-rays. The applied PSD method proved to be useful for n/gamma discrimination. A good resolving power of the method was achieved even without gating on neutron capture events. A comparison with a standard BC501A liquid scintillator was done to evaluate thermal neutrons detection efficiency in BC523A.
No takes yet. Share an insight, caveat, or question.
Świderski et al. (2008) studied this question.