Why the study?
Does rate responsive pacing mode (VVIR) improve exercise duration compared to fixed-rate demand mode (VVI) in patients with coronary artery disease and angina pectoris?
Does rate responsive pacing mode (VVIR) improve exercise duration compared to fixed-rate demand mode (VVI) in patients with coronary artery disease and angina pectoris?
In patients with CAD and angina, rate responsive pacing (VVIR) improves exercise capacity and heart rate response without worsening angina symptoms compared to fixed-rate pacing.
Answers are grounded in this paper and cite their sources.
VVIR pacing improves exercise capacity in CAD patients with angina; supports rate-responsive modes over fixed-rate VVI in practice.
The present study included 17 patients with angina pectoris and coronary artery disease in whom a rate responsive ventricular pacemaker (Medtronic Activitrax) had been implanted. All patients had an exclusively paced rhythm. Single blinded, random, cross-over treadmill tests in the rate responsive pacing mode (VVIR) and in the fixed-rate demand mode (VVI) were performed, with an interval of 4-6 weeks. Mean exercise duration increased by 25% during VVIR pacing. Maximal heart rate increased significantly during VVIR compared to VVI pacing (VVI = 74 +/- 2 bpm, VVIR = 116 +/- 8 bpm, P less than 0.001) as did the rate-pressure product (VVI = 10.850 +/- 1,124, VVIR = 16.628 +/- 2,110, P less than 0.001). Despite improved performance, the number of anginal attacks per week and the nitroglycerin consumption did not show a significant difference between the two pacing modes. It is concluded that rate responsive pacing is beneficial and safe in patients with angina pectoris and coronary artery disease.
No takes yet. Share an insight, caveat, or question.
Cock et al. (1989) studied this question.