Why the study?
What are the electrophysiological characteristics of the Medtronic 6991 transvenous atrial lead over four weeks post-insertion?
What are the electrophysiological characteristics of the Medtronic 6991 transvenous atrial lead over four weeks post-insertion?
Answers are grounded in this paper and cite their sources.
The Medtronic 6991 atrial transvenous lead shows a significant decrease in P wave amplitude during the first week and an increase in stimulation threshold over four weeks post-insertion.
Despite some reports on the use of the atrial transvenous endocardial lead, Medtronic 6991, there is still very little information on the electrophysiological properties of the lead. A lead extension wire was connected to the atrial lead, passed subcutaneously, and brought through the abdominal skin. P wave amplitude, threshold of stimulation, and electrode resistance were repeatedly determined through the atrial lead during four weeks after the electrode insertion in 17 patients. Mean P wave amplitude at insertion was 3-7 1 0 (mean SD) mV. It decreased significantly to a lowest mean level of 2-3 09 mV after one week. From that time there were only small variations. In the supine position and with normal breathing there was a spontaneous variation in the P wave amplitude of 15 per cent. P wave amplitude was only to a minor extent influenced by body position and maximum breathing movements. The threshold of stimulation was 1-7 0-7 V (impulse duration 0 5 mi) at the time of electrode insertion. It increased to 2 7 1 0 V after four weeks. The electrode resistance varied about 600 ohms.
No takes yet. Share an insight, caveat, or question.
Kruse et al. (1979) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: