Key result
Different implantation orientations of a bi-leaflet mechanical heart valve did not affect viscous shear stress distribution or total platelet activation.
Why the study?
Does implantation orientation of a bileaflet mechanical heart valve affect leaflet motion, shear stress, and platelet activation in an anatomic left ventricle-aorta model?
Does implantation orientation of a bileaflet mechanical heart valve affect leaflet motion, shear stress, and platelet activation in an anatomic left ventricle-aorta model?
Computational simulations suggest that the implantation orientation of a bileaflet mechanical heart valve does not significantly affect viscous shear stress or platelet activation in the ascending aorta.
Answers are grounded in this paper and cite their sources.
Orientation flexibility appears acceptable regarding shear; leaves open human clinical relevance.
We have performed three-dimensional high-resolution numerical simulations of a bi-leaflet mechanical heart valve implanted at different orientations in an anatomic left ventricle-aorta obtained from magnetic resonance imaging of a volunteer. The thoroughly validated overset curvilinear-immersed boundary fluid-structure interaction flow solver is used in which the aorta and left ventricle (LV) are discretized with boundary-conforming and nonconforming curvilinear grids, respectively. The motion of the left ventricle wall is prescribed based on a lumped parameter model while the motion of the leaflets is calculated using a strongly coupled fluid-structure interaction algorithm enhanced with Aitken convergence technique. We carried out simulations for three valve orientations, which differ from each other by 45 deg, and compared the leaflet motion and flow field for multiple cycles. Our results show reproducible and relatively symmetrical opening for all valve orientations. The presence of small-scale vortical structures after peak systole causes significant cycle-to-cycle variations in valve kinematics during the closing phase for all valve orientations. Furthermore, our results show that valve orientation does not have a significant effect on the distribution of viscous shear stress in the ascending aorta. Additionally, two different mathematical activation models including linear level of activation and Soares model are used to quantify the platelet activation in the ascending aorta. The results show that the valve orientation does not significantly affect (less than 8%) the total platelet activation in the ascending aorta.
No takes yet. Share an insight, caveat, or question.
Asadi et al. (2022) studied Mechanical heart valve implantation (n=1). Valve orientation vs. Different orientations (differing by 45 degrees) was evaluated on Leaflet motion, flow field, viscous shear stress, and platelet activation. Different implantation orientations of a bi-leaflet mechanical heart valve did not affect viscous shear stress distribution or total platelet activation.
Synapse has enriched 10 related papers. Consider them for comparative context: