Key result
Serial decremental rate inhibition identified 8% of VVI pacemaker patients as highly pacing dependent, while 26% developed an adequate rhythm but had symptoms with abrupt complete inhibition.
Why the study?
Does serial decremental rate inhibition safely and reliably identify pacemaker dependence in patients with VVI pacemakers compared to abrupt complete inhibition?
Observational (n=86)
Does serial decremental rate inhibition safely and reliably identify pacemaker dependence in patients with VVI pacemakers compared to abrupt complete inhibition?
Serial decremental rate inhibition is a reliable method to identify pacemaker-dependent patients and can prevent symptoms associated with abrupt pacing interruption.
Answers are grounded in this paper and cite their sources.
May identify pacing-dependent VVI patients without abrupt inhibition symptoms; leaves open prospective validation before practice change.
In a study of the concept of pacemaker dependence, 86 patients who had permanent ventricular demand (VVI) pacemakers implanted for more than six months and were in predominantly paced rhythm underwent repeated interruption of pacing both abruptly and after gradual reduction of paced rate. Non-invasive pace slowing was accomplished by timed inhibiting pulses applied to the chest wall. The results identified a small group (A) of patients, 8% of those tested, who consistently failed to develop an adequate escape rhythm and may thus be considered highly pacing dependent. A second group B, 26%, developed an adequate intrinsic rhythm with serial decremental rate inhibition (SDRI) but suffered symptoms with abrupt complete inhibition (ACI). The remaining 66%, group C, developed no symptoms in any test. Analysis of clinical pre-test data showed that syncope, atrioventricular block and a low rate prior to implant, and a long duration of pacing were associated with a higher degree of pacing dependence, but were not reliably predictive. The clinical significance and value of the results are discussed. We conclude that SDRI is a reliable test for the high degree of pacemaker dependence and can be helpful especially in planning a procedure which may involve interruption of pacing.
No takes yet. Share an insight, caveat, or question.
Crick et al. (1985) conducted an observational in Pacemaker dependence (n=86). Serial decremental rate inhibition vs. Abrupt complete inhibition was evaluated on Development of an adequate escape rhythm and symptoms upon pacing interruption. Serial decremental rate inhibition identified 8% of VVI pacemaker patients as highly pacing dependent, while 26% developed an adequate rhythm but had symptoms with abrupt complete inhibition.
Synapse has enriched 10 related papers. Consider them for comparative context: