Key result
Motion-mitigated MRI-guided STAR achieves ~99% dosimetric delivery accuracy using MLC tracking and cardiac gating.
Why the study?
Stereotactic arrhythmia radioablation targets for refractory VT are subject to both respiratory and rapid cardiac motion, presenting a unique challenge.
Does real-time cardiorespiratory motion-mitigated MRI-guided stereotactic arrhythmia radioablation improve dosimetric accuracy in phantom and in-silico models?
Does real-time cardiorespiratory motion-mitigated MRI-guided stereotactic arrhythmia radioablation improve dosimetric accuracy in phantom and in-silico models?
Effect estimate: 99% gamma pass rate
Real-time cardiorespiratory motion management for MR-linac-guided stereotactic arrhythmia radioablation is technically feasible and effectively mitigates cardiac motion-induced dosimetric uncertainties in experimental models.
Supports feasibility of motion-mitigated MRI-guided STAR in models; leaves open clinical validation in arrhythmia patients.
Objective. Stereotactic arrhythmia radioablation (STAR) is a novel, non-invasive treatment for refractory ventricular tachycardia (VT). The VT isthmus is subject to both respiratory and cardiac motion. Rapid cardiac motion presents a unique challenge. In this study, we provide first experimental evidence for real-time cardiorespiratory motion-mitigated MRI-guided STAR on the 1.5 T Unity MR-linac (Elekta AB, Stockholm, Sweden) aimed at simultaneously compensating cardiac and respiratory motions. Approach. A real-time cardiorespiratory motion-mitigated radiotherapy workflow was developed on the Unity MR-linac in research mode. A 15-beam intensity-modulated radiation therapy treatment plan (1 × 25 Gy) was created in Monaco v.5.40.01 (Elekta AB) for the Quasar MRI 4D phantom (ModusQA, London, ON). A film dosimetry insert was moved by combining either artificial (cos 4 , 70 bpm, 10 mm peak-to-peak) or subject-derived (59 average bpm, 15.3 mm peak-to-peak) cardiac motion with respiratory (sin, 12 bpm, 20 mm peak-to-peak) motion. A balanced 2D cine MRI sequence (13 Hz, field-of-view = 400 × 207 mm 2 , resolution = 3 × 3 × 15 mm 3 ) was developed to estimate cardiorespiratory motion. Cardiorespiratory motion was estimated by rigid registration and then deconvoluted into cardiac and respiratory components. For beam gating, the cardiac component was used, whereas the respiratory component was used for MLC-tracking. In-silico dose accumulation experiments were performed on three patient data sets to simulate the dosimetric effect of cardiac motion on VT targets. Main results. Experimentally, a duty cycle of 57% was achieved when simultaneously applying respiratory MLC-tracking and cardiac gating. Using film, excellent agreement was observed compared to a static reference delivery, resulting in a 1%/1 mm gamma pass rate of 99%. The end-to-end gating latency was 126 ms on the Unity MR-linac. Simulations showed that cardiac motion decreased the target's D98% dose between 0.1 and 1.3 Gy, with gating providing effective mitigation. Significance. Real-time MRI-guided cardiorespiratory motion management greatly reduces motion-induced dosimetric uncertainty and warrants further research and development for potential future use in STAR.
No takes yet. Share an insight, caveat, or question.
Akdag et al. (2022) studied Refractory ventricular tachycardia (phantom and in-silico models) (n=3). Real-time cardiorespiratory motion-mitigated MRI-guided STAR (tracking plus gating) vs. Static reference delivery was evaluated on 1%/1 mm gamma pass rate compared to static reference delivery (99% gamma pass rate). Real-time cardiorespiratory motion-mitigated MRI-guided STAR applying respiratory MLC-tracking and cardiac gating achieved a 57% duty cycle and a 1%/1 mm gamma pass rate of 99% compared to static reference delivery.