Key result
Device extraction for lead endocarditis using transoesophageal echocardiography guidance revealed a right atrial ghost, a cast-like structure representing fibrous or infective debris.
Case Report (n=1)
Highlights the occurrence of a right atrial 'ghost' (residual fibrous or infective debris) following ICD lead extraction for endocarditis, which should be documented due to its association with higher mortality.
Post-extraction right atrial ghosts warrant documentation given mortality link; case report leaves open prospective assessment of detection and outcomes.
A 64-year-old male required extraction of a dual chamber implantable cardioverter defibrillator (ICD) due to lead endocarditis. Blood cultures were positive for Staphylococcus lugdunensis. Transoesophageal echocardiography (TOE) demonstrated a 11.7 mm vegetation attached to the right ventricular (RV) lead (Fig. 1 and Video 1). No vegetation was identified on the right atrial (RA) lead. Device extraction took place using TOE guidance. It was uncomplicated, but at the end of the procedure a cast-like structure was seen in the right atrium thought to represent fibrous or infective debris which had been encasing one of the leads (Fig. 2 and Video 2). All lead insulation from the extracted leads was intact. Device lead endocarditis is an indication for complete removal of the pacemaker system and intra-procedural imaging such as TOE is strongly recommended during device extraction to assess for complications which may lead to haemodynamic instability (1). The presence of residual debris or ghosts in the RA or superior vena cava (SVC) following lead extraction has an incidence of 8–14% (2, 3), occurring more commonly with infected leads. There is no specific therapy when this is identified, but it has been associated with higher mortality (3). Its presence ought to be documented at the time of extraction. Open in a separate window Figure 1 Pre-extraction TOE, bi-caval view. A vegetation is seen attached to the RV lead as it enters from the SVC at the RA junction.
No takes yet. Share an insight, caveat, or question.
Pettemerides et al. (2019) conducted a case report in Lead endocarditis (n=1). Device extraction using TOE guidance was evaluated. Device extraction for lead endocarditis using transoesophageal echocardiography guidance revealed a right atrial ghost, a cast-like structure representing fibrous or infective debris.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: