Key result
TAVI valve resheathing shows no increase in mortality or short-term structural deterioration.
Why the study?
New transcatheter aortic valves allow resheathing and repositioning, but whether the resheath manoeuvre impairs patient outcomes and bioprosthesis durability after TAVI was unclear.
Does the resheathing maneuver of a self-expanding valve impair clinical outcomes or valve durability in patients undergoing TAVI?
Cohort (n=346)
No
Does the resheathing maneuver of a self-expanding valve impair clinical outcomes or valve durability in patients undergoing TAVI?
The resheath maneuver for self-expanding transcatheter aortic valves appears safe, demonstrating no adverse impact on medium-term survival or bioprosthetic valve durability.
Answers are grounded in this paper and cite their sources.
Resheathing appears safe short-term in this cohort; leaves open confirmation of durability in prospective randomized trials.
BACKGROUND: New transcatheter aortic valves were recently developed, enabling to resheath and reposition the prosthesis. The aim of the present study was to investigate whether the resheath manoeuvre did not impair the outcome of patients and the bioprosthesis durability after transcatheter aortic valve implantation (TAVI). METHODS AND RESULTS: On the 346 consecutive patients (84 ± 7 yrs-old, mean STS 6.7 ± 5%) undergoing a transfemoral TAVI in our institution since January 2008, 170 patients were implanted using a self-expanding valve (SEV). Among those, 39 (Group 1) required resheathing to achieve a successful implantation, while 131 did not require it (Group 2, N = 131). A balloon-expanding valve (BEV) was used in 176 patients (Group 3). Baseline characteristics were similar between groups. Device success was 98%, the rate of in-hospital death was 2%, and the number of procedural complications was similarly low, with no significant difference between groups. The follow-up was complete in 337 of 338 patients undergoing a successful TAVI (781 patients-year). Kaplan-Meier analysis showed that overall survival was 80 ± 2% and 42 ± 3% at 1 and 5 years respectively, with no difference between groups. On multivariate analysis, acute kidney injury, post-dilatation, pulmonary hypertension, porcelain aorta and STS score, but not resheath, were independant predictors of death after TAVI. The annual event rate of structural valve deterioration was 0.6% patients-year, and similar between groups. CONCLUSIONS: Our study shows that SEV resheath did not impair the procedural results, the outcome of patients nor the valve durability at short term after TAVI.
No takes yet. Share an insight, caveat, or question.
Kefer et al. (2020) conducted a cohort in Severe aortic stenosis undergoing transfemoral TAVI (n=346). Self-expanding valve resheathing vs. Self-expanding valve without resheathing was evaluated on Overall survival after TAVI (HR 0.90, 95% CI 0.42-1.95, p=0.79). Resheathing of self-expanding valves during TAVI did not increase the risk of death and was not associated with structural valve deterioration at short-term follow-up.
Synapse has enriched 10 related papers. Consider them for comparative context: