Key result
Sequential right-arm-first BP measurement potentially overestimates interarm systolic differences by up to ~7 mmHg.
Why the study?
Guidelines recommend initial blood pressure assessment on both arms, but because simultaneous measurement devices are often unavailable, the appropriateness of sequential measurements needed evaluation.
Does sequential blood pressure measurement accurately assess interarm blood pressure differences?
Cross-Sectional (n=7,838)
Does sequential blood pressure measurement accurately assess interarm blood pressure differences?
Answers are grounded in this paper and cite their sources.
Sequential blood pressure measurement may overestimate interarm differences and lead to more frequent right arm selection, suggesting simultaneous measurements may be necessary.
Guidelines recommend initial blood pressure assessment on both arms, with the higher-reading arm used for future measurements. As simultaneous measurement devices are often unavailable, we evaluated if sequential measurements are appropriate for assessing interarm differences. Blood pressure was measured in Shanghai using an automated system. Consecutive readings were taken with 1-min intervals, beginning with the right followed by the left arm, with a third reading if the difference between the first two exceeded 5 mmHg. In 7838 participants, right arm readings were consistently higher, with mean interarm systolic/diastolic differences of 4.59/1.01 mmHg if two readings on each arm, 6.24/1.45 mmHg if two readings on one and three readings on another arm, and 7.22/1.90 mmHg if three readings on each arm. This sequential measurement approach led to more frequent right arm selection for blood pressure measurement and potential overestimation of the interarm differences. Simultaneous measurements may be needed, but future studies are required to confirm this.
No takes yet. Share an insight, caveat, or question.
Sheng et al. (2026) conducted a cross-sectional in Blood pressure assessment (n=7,838). Sequential blood pressure measurements was evaluated on Mean interarm systolic/diastolic differences. Sequential blood pressure measurements starting with the right arm led to potential overestimation of interarm differences, with mean systolic differences ranging from 4.59 to 7.22 mmHg.