Key result
Advancing a buckled TEE probe into the stomach under fluoroscopy enables perforation-free removal in 3 cases.
Why the study?
Buckling of a transesophageal echocardiography probe is a rare but potentially life-threatening complication associated with a significant risk of esophageal perforation.
Does advancement of a buckled TEE probe into the stomach under fluoroscopic guidance allow for safe removal without esophageal perforation in patients with a buckled probe?
Case Report (n=3)
Does advancement of a buckled TEE probe into the stomach under fluoroscopic guidance allow for safe removal without esophageal perforation in patients with a buckled probe?
Advancement of a buckled TEE probe into the stomach under fluoroscopic guidance is a safe and effective method for straightening and removing the probe to prevent esophageal perforation.
Answers are grounded in this paper and cite their sources.
Multimodal imaging enabled safe resolution of buckled TEE probes in 3 cases; extends sparse reports but leaves optimal protocols open.
BACKGROUND: Buckling of a transesophageal echocardiography (TEE) probe is a rare but potentially life-threatening complication that carries a significant risk of esophageal perforation. CASE SUMMARY: We report 3 cases of buckled TEE probes that were promptly recognized, diagnosed, and managed using multimodal imaging. In each instance, the location of the buckled probe was confirmed with bedside imaging. Subsequently, the patient was taken to the interventional suite for advancement of the probe to the stomach under fluoroscopic guidance. Finally, the buckled TEE probe was straightened and removed successfully from the patient without difficulty. DISCUSSION: Early recognition of buckled TEE probes and employment of a structured, algorithmic management approach are essential to optimizing outcomes and preventing associated morbidity and mortality. TAKE HOME MESSAGES: Buckling of the TEE probe requires a high index of suspicion and prompt diagnosis to prevent catastrophic complications such as esophageal perforation. Advancement of the TEE probe in the stomach under fluoroscopic guidance facilitates straightening and removal of the probe without the risk of esophageal perforation.
No takes yet. Share an insight, caveat, or question.
Ho et al. (2026) conducted a case report in Buckled transesophageal echocardiography (TEE) probe (n=3). Fluoroscopic guidance for probe advancement and removal was evaluated on Successful straightening and removal of the buckled TEE probe without esophageal perforation. Advancement of a buckled TEE probe into the stomach under fluoroscopic guidance facilitated successful straightening and removal without esophageal perforation in 3 cases.
Synapse has enriched 10 related papers. Consider them for comparative context: