Why the study?
Probe malfunction during TEE is an underrecognized problem that can impact mitral regurgitation severity assessment.
Case Report (n=1)
No
Timely recognition of TEE probe color Doppler malfunction—indicated by discrepancies with TTE and continuous-wave Doppler, absent pulmonary vein color flow, and poor 2D image quality requiring high gain—prevents critical underestimation of mitral regurgitation severity prior to cardiac surgery.
Answers are grounded in this paper and cite their sources.
TEE probe malfunction may underestimate MR severity; leaves open whether routine probe verification reduces misclassification before cardiac surgery.
•Probe malfunction during TEE is an underrecognized problem.•MR severity assessment: host (e.g., BP), technical (e.g., gain), probe malfunction.•Clues for malfunction 1: disparity between TTE and TEE; CW and color Doppler.•Clues for malfunction 2: unexplained poor image quality; need for high gain.•Transducer temperature can affect image quality.
No takes yet. Share an insight, caveat, or question.
Gopalakrishnan et al. (2020) conducted a case report in Severe multivessel coronary artery disease and severe mitral regurgitation (n=1). Replacement of malfunctioning transesophageal echocardiography probe with a new transesophageal echocardiography probe vs. Initial malfunctioning transesophageal echocardiography probe was evaluated on Accurate assessment of mitral regurgitation severity via color Doppler.