Polar Verity Sense armband strongly correlates with H10 chest strap heart rates during adolescent HIIT.

View Full Paper
Ask AI
Bookmark
Share

Why the study?

The real-world validity of the Polar Verity Sense for heart rate monitoring in school settings among children and adolescents was yet to be determined.

Does the Polar Verity Sense armband accurately measure heart rate compared to the Polar H10 chest strap in adolescents during school-based high-intensity interval training?

Comparison

Polar Verity Sense vs Polar H10

Design

Validation study

Key result

The Polar Verity Sense armband demonstrated very strong within-participant correlation (r = 0.93, p < 0.001) with the Polar H10 chest strap for heart rate monitoring during HIIT in adolescents.

Authors

YLYong LiuABAlan R. BarkerMLMinghui Li

Discussion

Loading...

Member takes

Overview

May support armband use for adolescent HIIT monitoring; leaves open broader validation before clinical adoption.

Study Design

Type

Cross-Sectional (n=39)

Structured PICO

Does the Polar Verity Sense armband accurately measure heart rate compared to the Polar H10 chest strap in adolescents during school-based high-intensity interval training?

P
Population
39 adolescents (mean age 12.2 years, 36% female) participating in a 6-minute resistance-based high-intensity interval training session.
E
Exposure
Polar Verity Sense (PVS) armband for heart rate monitoring
C
Comparator
Polar H10 chest strap (criterion measure)
O
Outcome
Heart rate measurement validity (assessed via within-participant correlation, mean absolute error, mean absolute percentage error, and Bland-Altman plots)surrogate

Main Result

Effect estimate: r = 0.93

p-value: p=< .001

The Polar Verity Sense armband provides valid heart rate measurements compared to a traditional chest strap during high-intensity interval training in adolescents, offering a more practical monitoring tool.

Cite This Study

Liu et al. (2025) reported a cross-sectional. Polar Verity Sense (PVS) vs. Polar H10 was evaluated on within-participant correlation of heart rate (r = 0.93, p=< .001). The Polar Verity Sense armband demonstrated very strong within-participant correlation (r = 0.93, p < 0.001) with the Polar H10 chest strap for heart rate monitoring during HIIT in adolescents.

synapsesocial.com/papers/6a340f3d564550b2400eb01chttps://doi.org/10.1080/02640414.2025.2491163
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Measuring Heart Rate Variability Using Commercially Available Devices in Healthy Children: A Validity and Reliability Study2020 · 121 citations
  2. 2The Polar® OH1 Optical Heart Rate Sensor is Valid during Moderate-Vigorous Exercise2018 · 49 citations
  3. 3Validation of a Photoplethysmographic Heart Rate Monitor: Polar OH12019 · 57 citations
  4. 4Validation and Calibration for Embedding Rating of Perceived Exertion Into High-Intensity Interval Exercise in Adolescents: A Lab-Based Study2024 · 5 citations
  5. 5Validity of Wearable Activity Monitors during Cycling and Resistance Exercise2017 · 224 citations