Randomized trial demonstrates effective detection of trace Pb2+ in water, suggesting reliable monitoring for safety.
The presence of lead Pb 2+ in drinking water pose a significant risk to human health if its concentration exceeds permissible limits. Thus, the detection of Pb 2+ is essential for assessing water quality and facilitating its real‐time assessment. Here, S‐g‐C 3 N 4 is synthesized via a one‐pot thermal polymerization synthesis, modified with nafion (Nf) and named as S‐g‐C 3 N 4 /Nf/GCE. This modified material is employed for the selective detection of Pb 2+ in trace levels by differential pulse anodic stripping voltammetry (DPASV). The performance of the S‐g‐C 3 N 4 /Nf/GCE electrochemical sensor is optimized for operational parameters, namely deposition potential, time, and pH of electrolyte. The electrochemical sensor showed an excellent linear response towards Pb 2+ detection in the 5 nM–2.5 µM range. The sensor demonstrated a limit of detection (LOD) of 0.16 ± 0.004 nM and a sensitivity of 93.03 μA cm −2 μM −1 . The electrochemical sensor displayed very high selectivity towards Pb 2+ especially in the coexistence of other metal ions such as Na + , Ca 2+ , K + , Ni 2+ , Co 3+ , Cd 2+ , Cu 2+ , and Hg 2+ . The real‐time effectiveness of the electrochemical sensor is validated by employing it in tap water sample.
No takes yet. Share an insight, caveat, or question.
Ram et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: