Key result
During a 34-J ICD defibrillation shock, the highest body surface leakage energy was 46.8 mJ with subcutaneous contact and 34.6 mJ with skin surface contact, remaining at a safe level for bystanders.
Why the study?
Does defibrillation shock delivery by an ICD system cause significant electrical current leakage to the body surface in a canine model?
Does defibrillation shock delivery by an ICD system cause significant electrical current leakage to the body surface in a canine model?
Body surface leakage energy during an ICD defibrillation shock is minimal and unlikely to be harmful to a person touching the patient due to the short duration of the shock.
Answers are grounded in this paper and cite their sources.
Minimal leakage in canine ICD shocks suggests low bystander risk; leaves open human safety confirmation.
The leakage of electrical current to the body surface during defibrillation shock delivery by an implantable cardioverter-defibrillator (ICD) device (the Medtronic Jewel Plus PCD system) was evaluated in 5 dogs. The defibrillation shocks were delivered between the active-can implanted in the left subclavicular region and the endocardial lead placed in the right ventricle at the energy levels of 1, 2, 8, 12, 24 and 34 J. During each delivery, the electrical current leakage from the body surface was measured by electrodes connected to a circuit at 4 recording positions: (A) parallel-subcutaneous (the electrodes were fixed in the subcutaneous tissue of the left shoulder and the right lower chest, and the direction of the electrode vector was parallel to the direction of the defibrillation energy flow); (B) cross-subcutaneous (the electrodes were fixed in the subcutaneous tissue of the right shoulder and the left lower chest, and the vector of the electrodes was roughly perpendicular to the direction of the energy flow); (C) parallel-surface (the electrodes were fixed with ECG paste on the shaved skin surface at the left shoulder and the right lower chest); and (D) surface grounded (the electrodes were fixed on the shaved skin surface at the left shoulder and the left foot, which was grounded). The circuit resistance was set at a variable level (100-5,000 ohms) in accordance with the resistance measured through each canine body. Leakage energies were measured in 750 defibrillation shocks with each circuit resistance in 5 dogs. The leakage energy increased in accordance with the increase of the delivered energy and the decrease of the circuit resistance in all 4 recording positions. When the circuit resistance was set at 1,000 ohms, the leakage energy during shock delivery at 34 J was 32+/-17 mJ at position A, 5+/-9 mJ at B, 10+/-9 mJ at C, and 4+/-3 mJ at D (p=0.042). The peak current was highest at position A and was 87+/-22 mA with a circuit resistance of 1,000 ohms. The power of the leakage energy depended on the delivered energy and the impedance between the electrodes. The angle between the alignment of the recording electrodes and the direction of the energy flow was another important factor in determining the leakage energy. Although the peak current of the leakage energy reached the level of macro shock, the highest leakage energy from the body surface was considerably less because of the short duration of the shock delivery.
No takes yet. Share an insight, caveat, or question.
Niwano et al. (2001) studied Healthy canines (experimental model) (n=5). ICD defibrillation shock was evaluated on Leakage energy during 34 J shock delivery with 1,000 ohm circuit resistance across different positions (p=0.042). During a 34-J ICD defibrillation shock, the highest body surface leakage energy was 46.8 mJ with subcutaneous contact and 34.6 mJ with skin surface contact, remaining at a safe level for bystanders.