Key result
Wide bioprosthetic valve interstrut distance delays LVOT obstruction ~4.4-fold compared to narrow distance.
Why the study?
The anterior mitral leaflet contributes to LV function but is usually excised during bioprosthetic valve insertion; methods to safely retain it are needed.
Does a wide interstrut distance subtending the preserved anterior mitral leaflet mitigate left ventricular outflow tract obstruction during bioprosthetic mitral valve replacement?
Does a wide interstrut distance subtending the preserved anterior mitral leaflet mitigate left ventricular outflow tract obstruction during bioprosthetic mitral valve replacement?
In a sheep model, a wide bioprosthetic interstrut distance subtending a preserved anterior mitral leaflet mitigates the severity and precipitability of left ventricular outflow tract obstruction.
Answers are grounded in this paper and cite their sources.
May mitigate LVOTO risk with preserved AML in mitral replacement; hypothesis-generating in sheep model, human validation required.
Background: The anterior mitral leaflet (AML) contributes to left ventricular (LV) function but is normally excised at the time of a bioprosthetic valve insertion. This study aimed to investigate methods of safely retaining the AML at the time of mitral valve replacement. Methods: Five adult sheep (57 ± 3.8 kg) each underwent 3 insertions of a bioprosthetic mitral valve (asymmetric interstrut sectors) alternating the wide and narrow interstrut distance under the AML. Each insertion was performed on normothermic beating-heart cardiopulmonary bypass, with full retention of the native valve. After each valve insertion, continuous measurements of LV and aortic pressures were recorded with echocardiographic assessment of mitral valve function. If LV outflow tract obstruction (LVOTO) was not seen on the resumption of normal cardiac output, a bolus of adrenaline was given to precipitate it. Results: Thirteen of 15 valve insertions resulted in LVOTO caused by systolic anterior motion (SAM), independent of valve orientation. The wide interstrut distance subtending the AML was associated with a greater requirement for inotropic stress to precipitate an obstruction and was associated with late systolic rather than holosystolic obstruction. Conclusions: The predisposition to and nature of LVOTO due to SAM were associated with the bioprosthetic valve interstrut distance subtending the fully retained AML and may explain the survival differences in such patients. This model represents an effective method for research into prevention of LVOTO following mitral valve replacement with preservation of the native valve.
No takes yet. Share an insight, caveat, or question.
Brunel et al. (2021) studied Left ventricular outflow tract obstruction (LVOTO) following mitral valve replacement (n=9). Wide interstrut distance subtending the anterior mitral leaflet vs. Narrow interstrut distance subtending the anterior mitral leaflet was evaluated on Time to left ventricular outflow tract obstruction (LVOTO) (ß = 6.5, 95% CI 4.5-8.7, p=<.0001). A wide bioprosthetic valve interstrut distance subtending the fully retained anterior mitral leaflet significantly increased the time to left ventricular outflow tract obstruction compared to a narrow distance.
Synapse has enriched 10 related papers. Consider them for comparative context: