Key result
Transvenous pacemaker electrodes demonstrated a 1-year malfunction rate of 7.4%, with most occurring within the first 30 days and many showing clues of preventability.
Why the study?
What is the incidence and preventability of early malfunction of transvenous pacemaker electrodes?
Observational
Yes
What is the incidence and preventability of early malfunction of transvenous pacemaker electrodes?
Early transvenous pacemaker electrode malfunction occurs in 7.4% of cases at 1 year, with many cases being potentially preventable by recognizing specific intraoperative clues.
Answers are grounded in this paper and cite their sources.
May support intraoperative vigilance to reduce early transvenous lead malfunction; leaves open efficacy of prevention strategies in prospective studies.
A 3-year study by three medical centers has revealed a 1-year electrode malfunction rate of 7.4%; most malfunctions occurred within the first 30 days. The incidence of unavoidable early malfunction (3.2%) fell within the 5% standards suggested by the committee report of the Inter-Society Committee on Heart Diseases. Incidences of obscure cause (3.2%) may be difficult to identify prospectively and may be, to a certain extent, unavoidable. The majority of the malfunctions (4.2%) showed specific clues that indicated that they were preventable. Successful repositioning was achieved on the first attempt in 80.6% of the cases with malfunction, and only 0.7% required ultimate myocardial electrode implantation. The principal clues to potentially unsatisfactory positioning included the presence of a large right ventricle with or without tricuspid insufficiency, current thresholds greater than 0.5 mA and ST-segment deviations on the intracardiac electrogram of less than 2 mV. Electrode malfunction may be more common with bipolar than with unipolar electrodes; but significant differences in the incidence of malfunction among different unipolar electrodes were observed. These data indicate that further developments in transvenous electrode design are warranted.
No takes yet. Share an insight, caveat, or question.
Parsonnet et al. (1979) conducted an observational in Transvenous pacemaker electrode implantation. Transvenous pacemaker electrodes was evaluated on 1-year electrode malfunction rate. Transvenous pacemaker electrodes demonstrated a 1-year malfunction rate of 7.4%, with most occurring within the first 30 days and many showing clues of preventability.
Synapse has enriched 10 related papers. Consider them for comparative context: