Key result
MRI exposure in S-ICD patients is linked to a ~30% rate of alert tone loss.
Why the study?
Impaired perception of S-ICD warning tones or functional loss after MRI may compromise patient safety.
Does MRI exposure cause loss of S-ICD alert tone function, and what factors predict the inability to perceive the tone in S-ICD patients?
Observational (n=262)
Yes
Does MRI exposure cause loss of S-ICD alert tone function, and what factors predict the inability to perceive the tone in S-ICD patients?
MRI exposure significantly increases the risk of S-ICD alert tone malfunction, and older age or hypoacusis impairs tone perception, underscoring the importance of routine testing and telemetric surveillance.
Answers are grounded in this paper and cite their sources.
Functional alert tones appear largely preserved; leaves open perception gaps and needs prospective validation before guiding MRI protocols.
Background The subcutaneous implantable cardioverter-defibrillator (S-ICD) provides audible alert tones to indicate device or lead issues. Impaired perception of this warning tone or functional loss after magnetic resonance imaging (MRI) may compromise patient safety. Purpose This study investigated (1) factors associated with inability of patients to perceive the alert tone, and (2) the impact of MRI scans on post-scan tone function. Methods A total of 262 S-ICD patients from 3 European centers were prospectively included. The alert tone was demonstrated to all patients. Age, body mass index (BMI), sex, and diagnosed hypoacusis were analyzed as predictors of tone perception using multivariable logistic regression. In an additional analysis, a subgroup of 33 device patients who had previously undergone MRI scanning were evaluated for alert tone function loss, stratified by scan region (e.g. cardiac, cranial, spinal, abdominal). Differences across MRI regions were assessed using Fisher’s exact test and pairwise odds ratios. Results Alert tone testing confirmed functionality in 250/262 devices (95.4%). Of these 250 patients with a functional tone, 12 (4.8%) were unable to perceive it. Higher age and hypoacusis were independent predictors of failure to perceive the tone (both p < 0.05). In a separate analysis focusing on the technical integrity of the alert tone after MRI exposure, we evaluated the subgroup of patients who had undergone MRI scanning. Among 33 MRI-exposed devices, an objective loss of tone function post-scan occurred in 10 (30.3%). Only 0.9% of devices in patients without MRI did not have a functional tone; (p < 0.001). The incidence of tone loss varied between MRI regions without reaching the level of statistical significance (overall Fisher’s exact test, p = 0.81). No single MRI region showed a significantly higher risk compared with the reference region in pairwise analyses. Discussion/Conclusions In a large all-comer S-ICD cohort, 4.8% of patients are unable to hear the alert tone. Advanced age and pre-existing hypoacusis independently reduced the likelihood of perceiving the S-ICD alert tone. Undergoing an MRI had a significantly high risk of loss of alert tone function. These findings emphasize the importance of testing the audible alert and demonstrating it to patients. Telemetric surveillance should be considered in S-ICD management, especially in patients unable to perceive the alert tone.
No takes yet. Share an insight, caveat, or question.
Woermann et al. (2026) conducted an observational in Subcutaneous implantable cardioverter-defibrillator (S-ICD) (n=262). MRI scanning vs. No MRI scanning was evaluated on Objective loss of alert tone function (p=< 0.001). MRI exposure in S-ICD patients was associated with a higher rate of objective alert tone function loss compared to patients without MRI exposure.