Key result
Evolut PRO+ valve meets hydrodynamic criteria through 20 years of accelerated wear testing.
Why the study?
Does the Evolut PRO+ transcatheter heart valve maintain hydrodynamic performance through 30 years of accelerated wear testing?
Does the Evolut PRO+ transcatheter heart valve maintain hydrodynamic performance through 30 years of accelerated wear testing?
The Evolut PRO+ transcatheter heart valve demonstrates robust in vitro durability, meeting ISO hydrodynamic criteria for up to 20-30 simulated years.
Answers are grounded in this paper and cite their sources.
Supports potential extended durability of Evolut PRO+; leaves open translation to clinical longevity.
•First reported 30-year in vitro accelerated wear testing data for a THV.•All Evolut PRO+ valves met ISO hydrodynamic criteria through 20 years of testing.•At least one valve per size maintained performance through 30 years of testing.•Failures manifested in a consistent pattern, suggesting a shared mechanism.•Testing is ongoing until all valves fail.
No takes yet. Share an insight, caveat, or question.
Attizzani et al. (2026) studied this question. The Evolut PRO+ transcatheter heart valve met ISO hydrodynamic criteria through 20 years of accelerated wear testing, with at least one valve per size maintaining performance through 30 years.