Key result
Intravascular ultrasound grading of lead adherence correlated with transvenous lead extraction difficulty, with 86% of patients with high clinical difficulty scores showing low IVUS grades.
Why the study?
Clinical factors associated with intravascular lead adherence (ILA) are unreliable predictors, warranting accurate assessment because vascular injury during transvenous lead extraction is more likely in segments with higher ILA.
Does intravascular ultrasound (IVUS) accurately characterize intravascular lead adherence and correlate with transvenous lead extraction difficulty in patients undergoing lead extraction?
Observational
Does intravascular ultrasound (IVUS) accurately characterize intravascular lead adherence and correlate with transvenous lead extraction difficulty in patients undergoing lead extraction?
IVUS is a feasible imaging modality that can characterize intravascular lead adherence and correlates with extraction difficulty, potentially offering better prediction than clinical factors alone.
Answers are grounded in this paper and cite their sources.
IVUS may inform lead extraction planning; hypothesis-generating in this feasibility study and requires validation before practice change.
BACKGROUND: Clinical factors associated with development of intravascular lead adherence (ILA) are unreliable predictors. Because vascular injury in the superior vena cava-right atrium during transvenous lead extraction is more likely to occur in segments with higher degrees of ILA, reliable and accurate assessment of ILA is warranted. We hypothesized that intravascular ultrasound (IVUS) could accurately visualize and quantify ILA and degree of ILA correlates with transvenous lead extraction difficulty. METHODS: Serial imaging of leads occurred before transvenous lead extraction using IVUS. ILA areas were classified as high or low grade. Degree of extraction difficulty was assessed using 2 metrics and correlated with ILA grade. Lead extraction difficulty was calculated for each patient and compared with IVUS findings. Most patients with low lead extraction difficulty score had low ILA grades. Eighty-six percentage of patients with high lead extraction difficulty score had low IVUS grade, and the degree of transvenous lead extraction difficulty was similar to patients with low IVUS grades and lead extraction difficulty scores. CONCLUSIONS: IVUS is a feasible imaging modality that may be useful in characterizing ILA in the superior vena cava-right atrium region. An ILA grading system using imaging correlates with extraction difficulty. Most patients with clinical factors associated with higher extraction difficulty may exhibit lower ILA and extraction difficulty based on IVUS imaging. Graphic Abstract: A graphic abstract is available for this article.
No takes yet. Share an insight, caveat, or question.
Beaser et al. (2020) conducted an observational in Intravascular lead adherence in transvenous lead extraction. Intravascular ultrasound (IVUS) was evaluated on Degree of transvenous lead extraction difficulty (p=<0.001). Intravascular ultrasound grading of lead adherence correlated with transvenous lead extraction difficulty, with 86% of patients with high clinical difficulty scores showing low IVUS grades.