Key result
Sphere-9 lattice tip catheter for PVI linked to intraprocedural fibrin deposition and subsequent TIA.
Case Report (n=1)
This case report identifies a risk of fibrin clot formation and subsequent embolic events during pulsed-field ablation with the Sphere-9 lattice-tip catheter despite stringent anticoagulation.
Answers are grounded in this paper and cite their sources.
May signal device-specific embolic risk during pulsed-field ablation; leaves open need for systematic safety evaluation.
Introduction Pulsed‐field ablation with a lattice‐tip catheter has emerged as a promising technology for pulmonary vein isolation. We report a case of intraprocedural catheter malfunction with fibrin deposition potentially leading to a transient neurological deficit during ablation despite optimal anticoagulation. Case A 39‐year‐old man with symptomatic atrial fibrillation and tachycardia‐mediated cardiomyopathy underwent pulmonary vein isolation using the Affera mapping system and Sphere‐9 lattice tip catheter. During the procedure, despite uninterrupted oral anticoagulation and heparin given to meet manufacturer‐recommended activated clotting times, the catheter malfunctioned, and when removed, there was a white solid substance that was seen within the lattice tip. The procedure was stopped prematurely with incomplete pulmonary vein isolation. A pathologist confirmed the white substance was fibrin inside the lattice tip. The patient had symptoms of transient ischemic attack postoperatively and also suffered early recurrence of atrial fibrillation. Conclusion This case demonstrates fibrin deposition and possible embolic events using the Sphere‐9 lattice tip catheter despite stringent anticoagulation protocols. This highlights the importance of meticulous catheter inspection as well as provides a post‐market complication utilizing this catheter.
No takes yet. Share an insight, caveat, or question.
Roma et al. (2026) conducted a case report in Symptomatic atrial fibrillation and tachycardia-mediated cardiomyopathy (n=1). Sphere-9 lattice tip catheter for pulmonary vein isolation was evaluated on Catheter malfunction with fibrin deposition and embolic events. Use of the Sphere-9 lattice tip catheter for pulmonary vein isolation resulted in intraprocedural fibrin deposition within the catheter tip and a subsequent transient ischemic attack.
Synapse has enriched 10 related papers. Consider them for comparative context: