Key result
Lattice-tip pulsed-field ablation causes unintended right atrial lesions in ~100% of cases during PVI.
Why the study?
The impact of monopolar biphasic lattice-tip pulsed-field ablation on adjacent right atrial structures remains poorly characterized.
Does monopolar biphasic lattice-tip pulsed-field ablation create unintended right atrial lesions during pulmonary vein isolation for atrial fibrillation?
Case Report (n=4)
No
Does monopolar biphasic lattice-tip pulsed-field ablation create unintended right atrial lesions during pulmonary vein isolation for atrial fibrillation?
Monopolar biphasic lattice-tip pulsed-field ablation may create unintended right atrial lesions when the anatomical distance between right pulmonary veins and right atrial structures is short or catheter stability is compromised.
Answers are grounded in this paper and cite their sources.
Alerts to off-target right atrial lesions with monopolar biphasic lattice-tip PFA in close anatomy or instability; leaves open incidence and mitigation in larger studies.
Structured Abstract Background Monopolar biphasic pulsed-field ablation (PFA) delivered by a lattice-tip catheter may create deeper lesions than conventional bipolar systems. The impact on adjacent right atrial structures remains poorly characterized. Case summary We report four patients undergoing pulmonary vein isolation using a lattice-tip PFA catheter. Post-procedural electroanatomic mapping revealed unintended right atrial lesions involving the posterior right atrium, interatrial septum, or superior vena cava. In patients with short anatomical distances between right pulmonary veins and adjacent right atrial structures, lesions developed despite apparently successful pulmonary vein isolation. Catheter instability related to coughing under deep sedation appeared to contribute to broader lesion formation in two cases. Conclusion These cases suggest that monopolar biphasic lattice-tip PFA may create unintended lesions in adjacent right atrial tissue when anatomical proximity is short or catheter stability is compromised. Awareness of these mechanisms may help optimize procedural safety and lesion delivery.
No takes yet. Share an insight, caveat, or question.
Hirokami et al. (2026) conducted a case report in Atrial Fibrillation (Persistent and Paroxysmal) (n=4). Monopolar biphasic lattice-tip pulsed-field ablation (Sphere-9) was evaluated on Unintended right atrial lesion formation identified by post-procedural electroanatomic mapping. Monopolar biphasic lattice-tip pulsed-field ablation created unintended right atrial lesions in all 4 patients due to short anatomical proximity and catheter instability.
Synapse has enriched 10 related papers. Consider them for comparative context: