Key result
Exposure to a magnetic field from a magnetic stirring rod significantly decreased 12 of 14 platelet aggregation parameters induced by ADP or collagen compared to a neutral stirring rod.
Why the study?
Does a magnetic field from a magnetic stirrer affect platelet aggregation in platelet-rich plasma from healthy volunteers?
Does a magnetic field from a magnetic stirrer affect platelet aggregation in platelet-rich plasma from healthy volunteers?
Magnetic stir bars used in standard aggregometry significantly decrease platelet aggregation parameters, suggesting that standard measurement techniques may introduce artifacts.
Answers are grounded in this paper and cite their sources.
Magnetic stirrers may confound platelet aggregation readings; leaves open whether lab protocols need revision.
Platelets are subjected to extremely low frequency electromagnetic fields during standard aggregometry measurements owing to the use of a magnetic stir bar in the instrument. This study evaluates the effects of this magnetic field exposure on platelet aggregation by comparing the results obtained in a modified aggregometer. Blood samples from healthy volunteers were anticoagulated using citrate or heparin. Platelet-rich plasma (PRP) samples were prepared. A mechanical stirring device was attached to the aggregometer instead of the magnetic stir bar system. The PRP samples were stirred using a stirring rod tip that did not produce any magnetic fields in one channel of the aggregometer; in the other channel, a stirring rod carrying a small magnet at its tip was used. As a result, a magnetic field in the extremely low frequency range and in the amplitude range of 1.9-65 mT was applied to the platelets assigned to the channel where the magnetic stirring rod tip was used. Aggregation was induced using adenosine diphosphate (ADP), collagen, or epinephrine. The slopes, maximum aggregation values, and areas under the aggregation curves were compared between the magnetic and neutral stirring rod tip groups. For samples stirred with the magnetic stirring rod tip, a significant decrease was observed in 12 of the 14 parameters evaluated for aggregations induced with ADP or collagen compared to the neutral stirring rod tip, regardless of the method used for anticoagulation. This observation indicates that the magnetic stir bars used in standard aggregometry may significantly alter aggregation parameters and platelets may be possible targets of electromagnetic fields.
No takes yet. Share an insight, caveat, or question.
Sağdilek et al. (2013) studied Healthy volunteers. Magnetic stirring rod tip vs. Neutral stirring rod tip was evaluated on Platelet aggregation parameters (slopes, maximum aggregation values, and areas under the aggregation curves). Exposure to a magnetic field from a magnetic stirring rod significantly decreased 12 of 14 platelet aggregation parameters induced by ADP or collagen compared to a neutral stirring rod.