Key result
The subaxillary approach for cardiac device implantation resulted in 0 procedural complications, 0 deaths, and 0 infections over a mean follow-up of 65 months in 23 young adults.
Why the study?
Is the subaxillary approach for cardiac device implantation safe, feasible, and aesthetically satisfactory in young adults?
Observational (n=23)
No
Is the subaxillary approach for cardiac device implantation safe, feasible, and aesthetically satisfactory in young adults?
Answers are grounded in this paper and cite their sources.
The subaxillary approach for cardiac device implantation is a feasible and safe alternative that offers excellent aesthetic results for young adults.
OBJECTIVES: Pacemakers and implantable cardioverter defibrillators have become more common in younger individuals, owing to broader indications and technological advances. Our goal was to report our long-term experience of implanting cardiac devices in young adults via a subaxillary approach with a hidden single incision, to minimize the aesthetic impact of the device. METHODS: From 2003 to 2012, 23 patients underwent cardiac device implantation via a subaxillary approach while they were under general anaesthesia. The axillary vein was punctured under contrast-medium radiological guidance. The devices were positioned into the pocket under the pectoralis major muscle and over the fascia that envelopes the pectoralis minor muscle, through a lateral surgical access along the midaxillary line. RESULTS: Twenty-two single- and dual-chamber implantable cardioverter defibrillator and 1 dual-chamber pacemaker were implanted (mean age 32.2 ± 13.4 years). All procedures were successfully performed without complications; overall hospitalization after implantation was 3.3 days. During mean a follow-up of 65 ± 18 months, no deaths, infections or device migrations occurred. Two ventricular lead dislodgements requiring repositioning occurred within 1 month, and 1 defibrillation lead fracture occurred 5 years after implantation. All patients were extremely satisfied with the aesthetic result. A mild increase in the pacing threshold (0.88 ± 0.6 V at 0.4/0.5 ms at implantation vs 1.00 ± 0.3 V at 0.4/0.5 ms at 2-year follow-up, P = 0.063) and a statistically significant decrease in the pacing impedance (637 ± 161 ohm at implantation vs 499 ± 81 ohm at 2-year follow-up, P = 0.001) were observed, without any consequences. CONCLUSIONS: The subaxillary approach is a safe, feasible technique that requires a simple surgical procedure. The implantable cardioverter defibrillator functions properly at implantation and at long-term follow-up.
No takes yet. Share an insight, caveat, or question.
Bolzan et al. (2018) conducted an observational in Need for cardiac electronic device implantation (n=23). Subaxillary approach to cardiac electronic device implantation was evaluated on Procedural complications and long-term device function. The subaxillary approach for cardiac device implantation resulted in 0 procedural complications, 0 deaths, and 0 infections over a mean follow-up of 65 months in 23 young adults.
Synapse has enriched 10 related papers. Consider them for comparative context: