Key result
Extraction of a chronically implanted active fixation Attain Stability Quadripolar lead was performed successfully without specialist tools, despite significant adhesion at the fixation site.
Why the study?
No cases of extraction of a chronically implanted Attain Stability Quadripolar 4798 active fixation coronary sinus lead had been published.
Is extraction of a chronically implanted active fixation Attain Stability Quadripolar coronary sinus lead feasible and safe?
Case Report (n=1)
No
Is extraction of a chronically implanted active fixation Attain Stability Quadripolar coronary sinus lead feasible and safe?
This first reported case demonstrates that extraction of a chronically implanted active fixation quadripolar coronary sinus lead is feasible without specialized tools, though significant adhesions may require substantial traction.
Answers are grounded in this paper and cite their sources.
Single-case extraction success suggests feasibility without specialist tools; leaves open safety, reproducibility, and generalizability for chronic active fixation CS leads.
BACKGROUND: The Attain Stability Quadripolar 4798 lead is a relatively new quadripolar active fixation coronary sinus (CS) lead. No cases of extraction of a chronically implanted 4798 lead have been published to date. CASE SUMMARY: A 52-year-old man with a history of previous cardiac resynchronization therapy pacemaker (CRT-P) insertion and atrioventricular node ablation presented to our institution with a pocket infection 69 weeks after implantation. Directed intravenous antibiotic therapy was commenced and an extraction was performed the following day. Extraction of the right atrial and right ventricular leads was simple and achieved with gentle manual traction. Extraction of the CS lead was more difficult. Significant traction was required due to the formation of adhesions inside the CS but extraction of the lead was eventually successful without complication. Specialized extraction equipment was not required. A new contralateral CRT-P device was implanted, and the patient was discharged home. He remains well at 3 months of follow-up. DISCUSSION: We present the first case of extraction of a chronically implanted active fixation Attain Stability Quadripolar lead. Our experience demonstrates that this has been performed successfully without specialist tools and with preservation of the CS branch. Significant adhesion was noted at the site of active fixation, however. Potential difficulty with this lead's extraction should therefore be considered when contemplating its use.
No takes yet. Share an insight, caveat, or question.
Guella et al. (2020) conducted a case report in Pocket infection after CRT-P insertion (n=1). Extraction of Attain Stability Quadripolar 4798 lead was evaluated on Successful extraction of the lead. Extraction of a chronically implanted active fixation Attain Stability Quadripolar lead was performed successfully without specialist tools, despite significant adhesion at the fixation site.