Surgical extraction of an infected active fixation coronary sinus lead on a beating heart was greatly facilitated by the use of a tissue stabilizer.
Case Report (n=1)
The use of a tissue stabilizer can facilitate the surgical extraction of an entrapped active fixation coronary sinus lead on a beating heart.
Answers are grounded in this paper and cite their sources.
May facilitate beating-heart extraction in rare entrapments; leaves open need for prospective validation.
We present a 49-year-old man who required extraction of a cardiac resynchronization device due to lead infection. While 2 leads were successfully retrieved by a transvenous approach the third one with active fixation and entrapped into a coronary sinus side-branch, required surgical intervention. During the operation, performed on cardiopulmonary bypass on a beating heart, lead extraction was greatly facilitated by employment of a tissue stabilizer routinely used for off-pump coronary surgery.
No takes yet. Share an insight, caveat, or question.
A 2017 study conducted a case report in Cardiac device lead infection (n=1). Surgical lead extraction using a tissue stabilizer was evaluated on Successful lead extraction. Surgical extraction of an infected active fixation coronary sinus lead on a beating heart was greatly facilitated by the use of a tissue stabilizer.