Key result
Use of a large sizing balloon (40% vs 3% failure rate, P<0.0001) and the width of the left atrial opening were strong predictors of severe residual shunting after percutaneous PFO closure.
Cohort (n=301)
Absolute Event Rate: 40% vs 3%
p-value: p=<0.0001
Left atrial opening width and the use of a large sizing balloon are strong predictors of residual shunting after percutaneous PFO closure.
May guide balloon sizing to minimize residual shunt after PFO closure; hypothesis-generating pending prospective validation.
BACKGROUND: Percutaneous, mechanical closure of defects of the atrial septum fails to completely resolve shunting in up to 20% of cases. Little is known about the factors associated with device failure. METHODS: We measured the left atrial opening (X), right atrial opening (Z), tunnel length (Y), septum secundum, device-septum primum separation, and tunnel compressibility of the patent foramen ovale (PFO) in 301 patients with cryptogenic neurological events, PFO anatomy, and severe Valsalva shunting (Spencer Grade 5-5+). All patients then underwent percutaneous closure with the GORE®HELEX Septal Occluder device and were evaluated at 3 months for residual shunt by transcranial Doppler (TCD). RESULTS: Severe residual Valsalva shunt (TCD Grade 5-5+) was found at 3 months in 21 of 301 (7%) patients. X, Y, and Z were associated with failure with a high degree of statistical significance, whereas the width of the septum secundum, device-septum primum separation, and tunnel compressibility were not. An unanticipated finding was that 14 of 35 (40%) patients sized with a large balloon failed compared with 9 of 280 (3%) sized with a small balloon (P < 0.0001). In the multivariate logistic regression model, X (P = < 0.0001) and balloon size (P < 0.0001) were both strong predictors of failure. CONCLUSIONS: In an intracardiac echocardiography-defined PFO population, characterized by severe baseline Valsalva shunt and a high incidence of persistent (rest) shunting, association of six intracardiac measurements to closure device failure by multivariate logistic regression showed that the width of the left atrial opening was a strong predictor of residual shunting. An unanticipated finding was that use of a large sizing balloon was also strongly associated with failure.
No takes yet. Share an insight, caveat, or question.
Sorensen et al. (2012) conducted a cohort in Patent foramen ovale with cryptogenic neurological events (n=301). Large sizing balloon and left atrial opening width vs. Small sizing balloon and narrower left atrial opening was evaluated on Severe residual Valsalva shunt (TCD Grade 5-5+) at 3 months (p=<0.0001). Use of a large sizing balloon (40% vs 3% failure rate, P<0.0001) and the width of the left atrial opening were strong predictors of severe residual shunting after percutaneous PFO closure.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: