Key result
Higher ventricular pacing rates linked to ~14% greater cardiac output in bradycardic AF.
Why the study?
The optimal lower limit rate for VVI pacemakers in atrial fibrillation has not been established, and changes in cardiac output at different pacemaker settings remain incompletely understood.
Does increasing the lower rate limit of VVI pacemakers improve cardiac output in patients with bradycardic atrial fibrillation?
Observational (n=11)
No
Does increasing the lower rate limit of VVI pacemakers improve cardiac output in patients with bradycardic atrial fibrillation?
Increasing the VVI pacing rate from 60 to 70/min or higher in patients with bradycardic atrial fibrillation increases cardiac output, particularly in those with larger left ventricular end-systolic volumes.
Answers are grounded in this paper and cite their sources.
Higher VVI rates may raise cardiac output in bradycardic AF; leaves open optimal settings pending randomized trials.
Background While most VVI pacemakers in bradycardic patients are set to a low limit of 60/min, the optimal lower limit rate for VVI pacemakers in atrial fibrillation has not been established. Although an increase in heart rate within the normal range in the setting of a VVI pacemaker might be expected to lead to an increase in cardiac output with the shortening of the diastolic time, the changes in cardiac output at different pacemaker settings have not been fully clarified. Methods We included 11 patients with bradycardic atrial fibrillation who had VVI pacemakers implanted. Stroke volume was measured using the electrical cardiometry method (AESCULON Ⓡ mini; Osypka Medical) without pacing and at ventricular pacings of 60, 70, 80, and 90/min. Results Stroke volume decreased stepwise at ventricular pacing rates of 60, 70, 80, and 90/min (63.6 ± 11.2, 61.9 ± 10.6, 59.3 ± 12.2, and 57.5 ± 12.2 mL, p < .001), but cardiac output increased (3.81 ± 0.67, 4.33 ± 0.74, 4.74 ± 0.97, and 5.17 ± 1.09 L/min, p < .001). The rate of increase in cardiac output at a pacing rate of 70/min compared to 60/min correlated with left ventricular end‐systolic volume ( r = 0.711, p = .014). Conclusions Cardiac output increased at a pacing rate of 70 compared to 60 in bradycardic atrial fibrillation patients, and the rate of increase in cardiac output was greater in those with larger left ventricular end‐systolic volume.
No takes yet. Share an insight, caveat, or question.
Yokota et al. (2023) conducted an observational in Atrial fibrillation with bradycardia (n=11). Ventricular pacing at higher rates vs. Ventricular pacing at 60 beats/min was evaluated on Cardiac output (p=<0.001). Increasing ventricular pacing rates from 60 to 90 beats/min progressively increased cardiac output despite a stepwise decrease in stroke volume in patients with bradycardic atrial fibrillation.