Key result
Implantation of the Med-Zenith PT-Valve in patients with severe pulmonary regurgitation reduced right ventricular end diastolic volume index from 181.6 to 123.4 ml/m2 at 1 year.
Why the study?
Nearly two-thirds of patients with dilated RVOT were excluded from pulmonary valve transplantation due to a lack of size-matched valves.
Does Med-Zenith PT-Valve implantation improve hemodynamics and right ventricular volumes in patients with severe pulmonary regurgitation and dilated RVOT?
Does Med-Zenith PT-Valve implantation improve hemodynamics and right ventricular volumes in patients with severe pulmonary regurgitation and dilated RVOT?
The Med-Zenith PT-Valve demonstrates safety and feasibility for treating severe pulmonary regurgitation in patients with significantly enlarged RVOT, improving right ventricular volumes and exercise capacity at 1 year.
Answers are grounded in this paper and cite their sources.
May support PT-Valve use in dilated RVOT; leaves open confirmation via randomized trials.
Objective Nearly 2/3 of patients with dilated right ventricular outflow tract (RVOT) were excluded from pulmonary valves transplantation due to the lack of size-matched valves. Here, we explored the safety and efficacy of the Med-Zenith PT-Valve for the treatment of patients with severe pulmonary regurgitation. Methods 22 Patients with severe PR (grade 3+,4+) were enrolled based on the anatomical features of native RVOT and the valve design. The immediate, 3-months and 1-year post-procedural follow-up data were analyzed. Results The baseline mean systolic diameters in the distal main pulmonary artery (MPA), MPA sinus junction, MPA sinus, pulmonary annulus, RVOT aneurysm and muscular outlet measured with computed tomography were 33.6 ± 6.1, 34.0 ± 5.8, 37.9 ± 6.0, 32.4 ± 7.3, 41.9 ± 9.3, and 34.4 ± 8.0 mm, respectively. The PT-Valve landing zone was set within these levels. Successful valve implantations were achieved in all patients without noticeable device malposition, coronary artery compression, pulmonary branch obstruction or paravalvular leak during follow-ups. Post-procedural pulmonary artery diastolic pressure increased from 5.8 ± 3.1 to 11.3 ± 2.5 mmHg. In the 3-month and 1-year follow-up, the right ventricular end diastolic volume index reduced from the baseline 181.6 ± 29.0 to 143.7 ± 29.7 ml/m 2 and 123.4 ± 31.2 ml/m 2 , and the trans-pulmonary valve gradient decreased from 25.6 ± 22.2 to 10.64 ± 3.54 mmHg and 11.16 ± 3.0 mmHg, respectively. The 6-min walk distance increased from 416.6 ± 97.9 to 455.9 ± 64.6 m and 467.8 ± 61.2 m, respectively. Conclusion This clinical trial revealed favorable outcomes for the safety, efficacy and feasibility of the Med-Zenith PT-Valve in the treatment of severe PR with significantly enlarged RVOT.
No takes yet. Share an insight, caveat, or question.
Shang et al. (2022) studied Severe pulmonary regurgitation (n=22). Med-Zenith PT-Valve was evaluated on Right ventricular end diastolic volume index. Implantation of the Med-Zenith PT-Valve in patients with severe pulmonary regurgitation reduced right ventricular end diastolic volume index from 181.6 to 123.4 ml/m2 at 1 year.
Synapse has enriched 10 related papers. Consider them for comparative context: