Key result
End-stage renal disease in patients undergoing ICD lead extraction was associated with significantly worse 1-year survival compared to patients without ESRD (65.6% vs. 92.6%, P<0.001).
Why the study?
Does trans-venous ICD lead extraction in patients with end-stage renal disease have similar procedural outcomes and long-term survival compared to patients without ESRD?
Cohort (n=635)
No
Does trans-venous ICD lead extraction in patients with end-stage renal disease have similar procedural outcomes and long-term survival compared to patients without ESRD?
ICD lead extraction is procedurally safe and effective in ESRD patients, but long-term mortality remains substantial, largely driven by the underlying ESRD and ICD indication.
Answers are grounded in this paper and cite their sources.
Lead extraction appears safe in ESRD; leaves open whether it improves long-term survival in this high-risk group.
AIMS: End-stage renal disease (ESRD) increases the risk of implantable cardioverter-defibrillator (ICD) infection. We sought to define outcomes of lead extraction in patients with ESRD. METHODS AND RESULTS: Implantable cardioverter-defibrillator lead extractions at our institution from January 2006 to March 2014 were stratified by absence (Control-Ex, n = 465) or presence (ESRD-Ex, n = 43) of ESRD. Procedural outcomes and survival were determined by medical records review. Survival in the ESRD-Ex group was compared with a contemporaneous cohort with ESRD undergoing ICD lead implantation (ESRD-I, n = 127). Among extraction patients, those with ESRD were more likely to be extracted for infection (74.4% vs. 28.6%, P < 0.001). Extraction procedure success (Control-Ex: 97% vs. ESRD-Ex: 93%, P = 0.17) and procedural deaths (Control-Ex: 1.1% vs. ESRD-Ex: 2.3%, P = 0.413) were similar. Survival 1 year following extraction was worse in the ESRD-Ex group compared with the Control-Ex, with a survival rate of 65.6% vs. 92.6% (P < 0.001); these curves continued to diverge through year 3. One-year survival in the ESRD-Ex group was worse than among ESRD patients undergoing ICD implant (ESRD-I), but these curves converged and survival was similar by year 3. CONCLUSIONS: Implantable cardioverter-defibrillator lead extraction can be performed safely and effectively in patients with ESRD. However, despite high rates of procedural success, long-term mortality following extraction in ESRD patients is substantial. Much of the long-term mortality risk appears to be accounted for by the presence of ESRD and an indication for an ICD.
No takes yet. Share an insight, caveat, or question.
Shah et al. (2016) conducted a cohort in End-stage renal disease and implantable cardioverter-defibrillator lead extraction (n=635). End-stage renal disease (ESRD) vs. Absence of ESRD was evaluated on 1-year survival following extraction (p=<0.001). End-stage renal disease in patients undergoing ICD lead extraction was associated with significantly worse 1-year survival compared to patients without ESRD (65.6% vs. 92.6%, P<0.001).